5-Year Results From the AMPLATZER Amulet Left Atrial Appendage Occluder Randomized Controlled Trial
Key result: Amulet and Watchman devices showed similar 5-year outcomes for patients with atrial fibrillation
Randomized controlled trialNo clear difference
Is the Amulet occluder safe and effective compared to the Watchman 2.5 device?
Who
1878
Patients with nonvalvular atrial fibrillation at high stroke risk
Compared
7.4%Amulet occluder
7.1%Watchman device
Main outcome
Composite of ischemic stroke or systemic embolism
Result: 7.4% vs 7.1%; P = 0.851
Bottom line: The Amulet occluder and Watchman 2.5 devices showed similar 5-year clinical outcomes for stroke and death.
JACC · 2024-11-18 · Save PNG
Key points
- The trial randomized 1,878 patients to receive either the Amulet occluder or the Watchman 2.5 device to close the left atrial appendage.
- At the 5-year follow-up, 94.0% of Amulet patients and 90.9% of Watchman patients were free of oral anticoagulation, which is blood-thinning medication.
- Clinical outcomes were similar between groups, including ischemic stroke or systemic embolism (7.4% vs 7.1%) and major bleeding (20.1% vs 20.0%).
Discussion prompts
- Would these patients look like yours? Who was left out?
- Is the main outcome one that matters to patients?
- Is the difference large enough to change what you do?
Cite: Lakkireddy D, Ellis CR, Thaler D, et al. 5-Year Results From the AMPLATZER Amulet Left Atrial Appendage Occluder Randomized Controlled Trial. J Am Coll Cardiol. 2025;85(11):1141-1153. doi:10.1016/j.jacc.2024.10.101