Randomized trialThe Lancet Digital Health · Sep 24, 2021

Fully closed-loop insulin system was not non-inferior to hybrid systems requiring meal input

A fully artificial pancreas versus a hybrid artificial pancreas for type 1 diabetes: a single-centre, open-label, randomised controlled, crossover, non-inferiority trial

Tsoukas MA, Majdpour D, Yale JF, Fathi AE, Garfield N, Rutkowski J, et al
Digital Health & AIEndocrinology & DiabetesGastroenterology & HepatologyAdult
Randomised, open-label, crossover, non-inferiority trialNo clear difference
Is a fully closed-loop system non-inferior to a hybrid closed-loop system?
Who
24
Adults with type 1 diabetes on insulin pump therapy · McGill University Health Centre, Montreal, Canada
Compared
74.3%Fully closed-loop system (IQR 61.5-82.8)
78.1%Hybrid closed-loop system (66.3-87.5)
Main outcome
Percentage of time spent within the glucose target range
Result: paired difference 2.6%, 95% CI -2.4 to 12.2
Bottom line: The fully closed-loop system was not non-inferior to the hybrid closed-loop system for time in target range.
Limitation: Outpatient studies comparing systems with patient-estimated carbohydrate counting are warranted.
Funded by: Diabetes Canada
Lancet Digit Health · 2021-09-24Download image

Clinical question: Is a fully closed-loop system non-inferior to a hybrid closed-loop system?

Why it matters: It is unknown if an automated insulin delivery system that does not require meal input is as effective as hybrid systems that require carbohydrate counting.

Key points

Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.

Was this summary helpful?

Read the full paper free at the publisher ↗⬇ Slide🖨 Handout
Top Digital Health & AI studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Abstract

Background For people with type 1 diabetes, there is currently no automated insulin delivery system that does not require meal input. We aimed to assess the efficacy of a novel faster-acting insulin aspart (Fiasp) plus pramlintide fully closed-loop system that does not require meal input. Methods In this open-label, randomised controlled, crossover, non-inferiority trial we compared the Fiasp (Novo Nordisk, Bagsværd, Denmark) plus pramlintide closed-loop system with no meal input (fully artificial pancreas) and the Fiasp-alone closed-loop system with precise carbohydrate counting (hybrid artificial pancreas). Adults (≥18 years) who had a clinical diagnosis of type 1 diabetes for at least 12 months, had glycated haemoglobin 12% or lower, and had been on insulin pump therapy for at least 6 months were enrolled at McGill University Health Centre, Montreal, QC, Canada. The Fiasp plus pramlintide fully closed-loop system delivered pramlintide in a basal-bolus manner with a fixed ratio of 10 μg:U relative to insulin. A research staff member counted the carbohydrate content of meals to input in the hybrid closed-loop system. Participants completed the two full-day crossover interventions in a random order allocated by a computer-generated code implementing a blocked randomisation (block size of four). The primary outcome was the percentage of time spent within the glucose target range (3·9-10·0 mmol/L), with a 6% non-inferiority margin, assessed in all participants who completed both interventions. This trial is registered with ClinicalTrials.gov, NCT03800875. Findings Between Feb 8, 2019, and Sept 19, 2020, we enrolled 28 adults, of whom 24 completed both interventions and were included in analyses. The percentage of time spent in the target range was 74·3% (IQR 61·5-82·8) with the fully closed-loop system versus 78·1% (66·3-87·5) with the hybrid Fiasp-alone closed-loop system (paired difference 2·6%, 95% CI -2·4 to 12·2; non-inferiority p=0·28). Eight (33%) participants had at least one hypoglycaemia event (<3·3 mmol/L) with the fully closed-loop system compared with 14 (58%) participants with the hybrid closed-loop system (2200-2200 h). Non-mild nausea was reported by three (13%) participants and non-mild bloating by one (4%) participant with the fully closed-loop system compared with zero participants with the hybrid closed-loop system. Interpretation The Fiasp plus pramlintide fully closed-loop system was not non-inferior to the Fiasp-alone hybrid closed-loop system for the overall percentage of time in the glucose target range. However, participants still spent a high percentage of time within the target range with the fully-closed loop system. Outpatient studies comparing the fully closed-loop hybrid systems with patient-estimated, rather than precise, carbohydrate counting are warranted. Funding Diabetes Canada.

Abstract reproduced under the article’s CC BY licence.

Subject terms

Adult; Blood Glucose; Canada; Cross-Over Studies; Diabetes Mellitus, Type 1; Drug Combinations; Female; Glycated Hemoglobin; Humans; Hypoglycemia; Hypoglycemic Agents; Insulin

DOI: 10.1016/S2589-7500(21)00139-4 · PubMed: 34580055 · Registered: NCT03800875

Cite this article

Tsoukas MA, Majdpour D, Yale JF, Fathi AE, Garfield N, Rutkowski J. A fully artificial pancreas versus a hybrid artificial pancreas for type 1 diabetes: a single-centre, open-label, randomised controlled, crossover, non-inferiority trial. Lancet Digit Health. 2021. doi:10.1016/S2589-7500(21)00139-4

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Randomized trial · Diabetes Obes Metab · Nov 2, 2021 · Endocrinology & Diabetes

Insulin degludec was associated with fewer nocturnal symptomatic hypoglycaemia episodes than insulin glargine U100

Randomized trial · J Clin Endocrinol Metab · Jul 15, 2022 · Endocrinology & Diabetes

Liraglutide treatment did not change glucose production or disposal during hypoglycemia in type 1 diabetes

Randomized trial · Diabetes Care · Jul 17, 2023 · Endocrinology & Diabetes

Insulin degludec and liraglutide showed similar blood sugar reduction compared to basal-bolus insulin therapy

Meta-analysis · Diabetes Obes Metab · Apr 2, 2025 · Endocrinology & Diabetes

iGlarLixi was associated with greater blood glucose reduction and weight benefit than IDegAsp

Randomized trial · Diabetes Obes Metab · Jul 25, 2022 · Endocrinology & Diabetes

iGlarLixi showed greater blood sugar reduction and weight loss than insulin glargine in Chinese adults