A Novel Nonretinal Fixation System for Epiretinal Prostheses: Preclinical Evaluation of Implant Stability and Long-Term Ocular Safety
Raz-Prag D, Rigler D, Mendelewicz R, Bez M, Stalmans P, Weinberger D, et al
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Purpose
To evaluate a novel sulcus-based, tack-free fixation strategy for the NanoRetina 600 (NR600) epiretinal prosthesis and assess surgical feasibility, implant stability, and long-term ocular safety in a porcine model. Design
Preclinical experimental animal study. Subjects
Twenty-one pigs undergoing unilateral intraocular implantation or device-related testing across surgical feasibility, implant delivery system (IDS), and in vivo activation cohorts. Methods
The wireless NR600 epiretinal prosthesis with penetrating microelectrodes and a self-expanding nitinol sulcus-fixation system was implanted in pigs with and without an IDS. Clinical examinations, ultrasonography, pachymetry, fluoroscopy, gross pathology, histopathology, and impedance testing were performed over 2 to 12 weeks. Main outcome measures
Procedural feasibility, adverse events, implant position and stability, electrode-retina apposition, anterior segment tolerance, and ocular tissue response. Results
Implantation was completed in all eyes. Most intraoperative complications were mild or moderate, procedure-related, and resolved without preventing successful implantation. Two eyes were excluded early because of serious postoperative events: one developed a complete retinal detachment in association with posterior dislocation of a sulcus-placed intraocular lens, and the other developed elevated intraocular pressure, pain, corneal opacity, and hyphema followed by phthisis bulbi. In the remaining eyes, the device remained stably positioned with minimal ocular tissue response, preserved retinal architecture beneath the array, and no sustained adverse effect on the anterior segment. Histopathology showed generally close implant-retina apposition with minimal inflammation or fibrosis, although electrode penetration was absent in one eye and partial in one eye; impedance measurements remained within the expected range, supporting electrical continuity and electrode-tissue contact in tested eyes. Conclusions
A sulcus-based, nonretinal fixation strategy enabled stable placement of the NR600 epiretinal prosthesis in pigs without retinal tacks and was associated with acceptable long-term ocular safety. This approach may simplify implantation and support further translational development. Financial disclosures
Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
DOI: 10.1016/j.xops.2026.101420 · PubMed: 43034847
Spotted a mistake? Suggest a correction
More in Ophthalmology
ResearchPediatricAm J Ophthalmol · Nov 20, 2026 · Ophthalmology
Association of allergic conjunctivitis severity with meibomian gland abnormalities in children: a cross-sectional study
Low-Concentration Atropine Eye Drops in Children With Myopia and Intermittent Exotropia: The AMIXT2 Randomized Clinical Trial
ResearchAm J Ophthalmol · Nov 19, 2026 · Ophthalmology
Smoke, Not Smoking, as a Possible Risk Factor Triggering Conversion in Leber's Hereditary Optic Neuropathy: A Retrospective Case Series
ResearchOphthalmology · Nov 18, 2026 · Ophthalmology
Trends and Variation in Uveal Melanoma Management in the National Cancer Database, 2004-2023
ResearchAm J Ophthalmol · Nov 18, 2026 · Ophthalmology
Developing Static Perimetry Thresholds to Assess Visual Field Loss Severity in Patients with Retinitis Pigmentosa