Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study

Duchatelet C, Segher K, Tackaert T, Vanwulpen M, Ingels B, Muylaert D, et al
Emergency MedicineCritical CareCardiologyAdult
Read the full paper free at PubMed Central ↗

Abstract

Aim This study aimed to evaluate the accuracy of two commercially available ventilation feedback devices (VFD) in measuring ventilation rate and tidal volume during simulated cardiopulmonary resuscitation (CPR) with continuous chest compressions and asynchronous ventilations in the presence of reversed airflow. Methods Prospective, simulation-based, non-randomized cross-over study. Twenty emergency department nurse practitioners performed manual ventilation on an intubated high-fidelity manikin during two 1-minute trials (one per device), targeting 10 breaths/min during continuous chest compressions (100-120/min). Ventilation parameters displayed by EOLifeX® and AccuVent® were compared with measurements from a high-fidelity pneumotachograph (HFRP). Data is presented as mean (standard deviation, min-max range). CI = confidence interval. Results EOLifeX®: From HFRP flow data, 254 ventilations were identified (rate 12/min [1; 8-14], tidal volume 544 mL [32; 482-604]). All showed reversed airflow with an average of 49 mL [13; 19-72]. VFD displayed values overestimated ventilation rate and underestimated tidal volume (111/min [5; 99-121] and 104 mL [14; 68-240]; biases +99 breaths/min, 95% CI 96.6-101.4, and -440 mL, 95% CI -456.3 to -423.7). AccuVent®: From HFRP flow data, a total of 222 ventilations were identified (rate 12/min [1; 8-13], tidal volume 532 mL [35; 460-614]). All showed reversed airflow with an average of 56 mL [15; 22-78]. VFD displayed values were similar for rate (12/min [1; 8-13]) but underestimated tidal volume (360 mL [24; 292-398]; bias -172 mL, 95% CI -191.9 to -152.1). Conclusion During simulated CPR with continuous compressions and asynchronous ventilations, VFDs demonstrated clinically relevant inaccuracies. Validation under realistic resuscitation conditions are needed.

DOI: 10.1016/j.resplu.2026.101475 · PubMed: 42761754

Cite this article

Duchatelet C, Segher K, Tackaert T, Vanwulpen M, Ingels B, Muylaert D. Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study. Resusc Plus. 2026. doi:10.1016/j.resplu.2026.101475

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

More in Emergency Medicine

J Med Internet Res · Nov 23, 2026 · Meta-analysis · Digital Health & AI, Emergency Medicine

Large language models mainly used for documentation in emergency departments

BMJ Med · Nov 18, 2026 · Research · Emergency Medicine

Direct admission to level 1 trauma centre was linked to lower 30 day mortality

Am J Emerg Med · Nov 13, 2026 · Research · Emergency Medicine, Hematology

ED red blood cell transfusion linked to longer hospital stays in severe anemia

J Med Internet Res · Nov 13, 2026 · Research · Digital Health & AI, Emergency Medicine

Physicians often intend to act on false information in AI-generated abstracts

Am J Emerg Med · Nov 12, 2026 · Research · Emergency Medicine, Neurology

Serum lactate levels differed across neurologic emergencies in emergency patients