Annuloplasty Ring Size and Long-Term Durability After Degenerative Mitral Valve Repair
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Background
Annuloplasty ring sizing has an important role in surgical mitral valve (MV) repair for degenerative mitral regurgitation (MR). Larger rings are generally preferred for degenerative disease as they confer optimal hemodynamics, reduce systolic anterior motion, and restore annular geometry. Methods
We analyzed a retrospective, single-surgeon cohort of 503 patients who underwent MV repair for MR from 2008 to 2024. The primary end point was MV repair failure, defined as reintervention, recurrent MR, or mitral stenosis. Patients were stratified by annuloplasty ring size quartiles. Kaplan-Meier survival and multivariable Cox proportional hazards models were used, incorporating a time-dependent spline interaction to capture long-term effects. Results
Over a median follow-up of 5.3 years (interquartile range: 2.8-8.2), larger annuloplasty ring size was significantly associated with higher rates of late repair failure. Failure rates by ring size quartile were ≤32 mm (2.0%), 33 to 34 mm (3.2%), 35 to 36 mm (5.7%), and >36 mm (7.3%). Ring size remained an independent predictor of MV repair failure (hazard ratio, 1.06; 95% CI, 1.01-1.10; P 34 mm was identified as the optimal cut point for predicting long-term failure. Conclusions
Larger annuloplasty ring size is independently associated with late MV repair failure. These findings support more physiologic sizing strategies to optimize long-term durability after MV repair for degenerative MR.
Spotted a mistake? Suggest a correction