Scar tissue at the lead site was associated with higher pacing thresholds in His-bundle pacing
Cardiovascular magnetic resonance findings in His-bundle pacing: Insights to pacing threshold rise
Why it matters: The mechanisms that cause pacing thresholds to rise despite an intact lead in His-bundle pacing are not well understood.
- Researchers studied 16 patients with His-bundle pacing systems, including 9 patients who experienced a significant pacing threshold increase of ≥1.0V/1.0 ms.
- The presence of late gadolinium enhancement, which indicates scar tissue, at the lead implantation site was associated with increased pacing thresholds.
- All magnetic resonance imaging scans were evaluable without visual artifacts and no periprocedural safety issues occurred during the study.
Limitation the authors note: The authors state that the small sample size means these exploratory findings are hypothesis-generating.
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.