Characterization of external defibrillator output and its impact on defibrillation protection of medical equipment

Zhang D, Deng J, Yin Y
Emergency MedicineAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Background Defibrillators constitute a critical component of emergency care for ventricular fibrillation and pulseless ventricular tachycardia. Advances in defibrillation technology point to the potential need for reassessing current requirements for defibrillation protection and associated resuscitation guidelines. Objective This study aimed to establish reference information for future directions in defibrillator requirements, defibrillation protection requirements, and resuscitation guidelines based on the parameters of commercially available defibrillators. Methods Defibrillator brands were identified from the registration records of the China National Medical Products Administration, the U.S. Food and Drug Administration, and the European Database on Medical Devices. Key defibrillation parameters were obtained from publicly available manufacturer documentation. Results A total of 31 manufacturers and 115 external defibrillators were identified. All studied devices employ biphasic waveforms, with biphasic truncated exponential being the most common, appearing in products from 24 of 31 manufacturers (77.4%). Except for pulsed biphasic devices manufactured by Schiller, maximum output voltages are below 3 kV. Maximum energy limits vary across devices, with some models retaining an upper limit of 360 J. Conclusion Incorporating biphasic waveforms into defibrillation protection tests may be appropriate. Adjustments to the requirements and guidelines of defibrillation voltage or energy could be considered, depending on market trends and future studies on defibrillation efficacy.

DOI: 10.1016/j.resplu.2026.101289 · PubMed: 41909181

Cite this article

Zhang D, Deng J, Yin Y. Characterization of external defibrillator output and its impact on defibrillation protection of medical equipment. Resusc Plus. 2026. doi:10.1016/j.resplu.2026.101289

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top studies every Monday

The week’s key trials and guidelines in your specialties, plus any topics you follow. By email, free.

Get the weekly digest

More in Emergency Medicine

Meta-analysis · J Med Internet Res · Nov 23, 2026 · Digital Health & AI, Emergency Medicine

Large language models mainly used for documentation in emergency departments

Research · BMJ Med · Nov 18, 2026 · Emergency Medicine

Direct admission to level 1 trauma centre was linked to lower 30 day mortality

Research · Am J Emerg Med · Nov 13, 2026 · Emergency Medicine, Hematology

Emergency department blood transfusions were associated with longer stays and higher rates of return

Research · J Med Internet Res · Nov 13, 2026 · Digital Health & AI, Emergency Medicine

Physicians often intend to act on false information in AI-generated abstracts

Research · Am J Emerg Med · Nov 12, 2026 · Emergency Medicine, Neurology

Serum lactate levels were associated with different types of neurologic emergencies in emergency department patients