ResearchClinical Journal of the American Society of Nephrology · Dec 5, 2025

Clinical Evaluation of a Disposable Test Strip for Potassium Level Measurement in Patients Treated with Maintenance Hemodialysis

Park CH, Kim Y, Kim C, Kim C, Kim S, Lee K, et al
NephrologyAdult
Read the full paper free at PubMed Central ↗
Top Nephrology studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Abstract

Key points Potassium level test strip showed excellent performance with accuracy and reliability in experimental setting with standard electrolyte solutions. Potassium levels obtained with test strips yielded satisfactory agreement with those measured by central laboratory equipment. Background Hyperkalemia is a common electrolyte imbalance leading to an increased risk of serious cardiac dysrhythmias and mortality in patients undergoing hemodialysis. A convenient point-of-care testing (POCT) of potassium levels is considered to improve patient outcomes. However, POCT for potassium level using capillary blood with satisfactory performance is not available. Methods We tested the performance of disposable test strips for potassium level using various concentrations of potassium solutions with or without interfering electrolytes in an experimental setting. Subsequently, we examined the agreement between potassium levels measured by disposable test strips and those measured by central laboratory equipment using various types of blood samples, including capillary blood obtained by finger prick, in 40 patients undergoing maintenance hemodialysis. In the analysis, Passing-Bablok regression and Bland-Altman analysis were used. Results Potassium concentrations measured by disposable test strips showed a high degree of agreement with potassium levels achieved using inductively coupled plasma atomic emission spectroscopy. The coefficient of variations for disposable test strips were <5% across potassium concentrations ranging from 2 to 9 mM. Furthermore, potassium levels measured by disposable strips from capillary blood demonstrated a high degree of agreement with potassium levels obtained by central laboratory equipment using serum, showing the slope achieved from Passing-Bablok regression was 1.04 (95% confidence interval, 0.97 to 1.12). Additional analysis with various blood samples also showed similar results. Conclusions POCT for potassium level using the disposable test strip could be used for self-monitoring of blood potassium levels in clinical practice.

Subject terms

Humans; Renal Dialysis; Potassium; Middle Aged; Male; Female; Aged; Reagent Strips; Point-of-Care Testing; Hyperkalemia; Reproducibility of Results; Disposable Equipment

DOI: 10.2215/CJN.0000000934 · PubMed: 41348486

Cite this article

Park CH, Kim Y, Kim C, et al. Clinical Evaluation of a Disposable Test Strip for Potassium Level Measurement in Patients Treated with Maintenance Hemodialysis. CJASN. 2025. doi:10.2215/CJN.0000000934

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Randomized trial · Kidney Int · Jul 4, 2025 · Nephrology, Cardiology

Sodium zirconium cyclosilicate did not change cardiovascular outcomes in patients receiving hemodialysis

Randomized trial · Kidney Int · Oct 26, 2024 · Nephrology, Cardiology

Combining higher potassium dialysate with sodium zirconium cyclosilicate reduced heart rhythm disturbances in hemodialysis patients

Randomized trial · J Clin Endocrinol Metab · Mar 15, 2023 · Endocrinology & Diabetes, Nephrology

Low-dose spironolactone reduced urine albumin levels without causing high potassium in diabetes patients

Randomized trial · Eur Heart J · Apr 22, 2025 · Cardiology, Nephrology

New risk score identifies patients at risk of high potassium in kidney disease and diabetes

Randomized trial · JASN · Nov 3, 2021 · Nephrology

Finerenone increased risk of high potassium levels in kidney disease patients