ResearchThe Laryngoscope · Jun 4, 2026
Cochlear Implants and Electrocautery: A Live Animal Model
Jeyakumar A, Lu Q, Johnson R, Mitchell R, Brickman T
Study type: Laboratory or animal research: results may not apply to people yet.
Abstract
Objectives
The use of monopolar cautery is regarded as a contraindication in cochlear implant (CI) recipients, despite a large amount of contradicting data. The objectives are as follows: (1) Evaluate the effects of high-powered electrocautery on CI devices. (2) Record any high-powered electrocautery induced temperature changes. Methods
After IACUC approval, 12 manufacturer-donated CIs were implanted in two anesthetized pigs. All devices underwent pre-study integrity checks; six had onsite impedance testing, while six received post-cautery analysis. Fifty watts coagulation was applied for 20 min at sites 1-15 cm from the CI. Bluetooth thermometers monitored three sites of temperatures. Devices were explanted for analysis. Stata 17.0 was used for statistical analysis. A p-value Results
The mean temperatures at the three sites (a*, b*, c*) was 32.9°C-38.8°C. Maximum absolute temperature changes averaged 4.1°C (95% CI: 1.4-6.8) for scalp measurements and 2.9°C (95% CI: 1.0-4.9) for round window measurements. Neither time (p = 0.09) nor distance (p = 0.64) generated significant scalp temperature changes. Devices were stratified by proximity. Mean maximum temperature changes showed a gradient by distance. Differences in device proximity were not statistically significant. Postimplantation device analysis was performed on six devices, and all passed the manufacturer in-depth device analysis. Conclusions
This controlled experimental evidence, demonstrating 100% device integrity across all tested conditions, challenges current manufacturer contraindications and supports evidence-based guideline revision. Level of evidence
N/A.
Topics
Animals; Electrocoagulation; Cochlear Implants; Swine; Models, Animal; Cochlear Implantation
DOI: 10.1002/lary.70649 · PubMed: 42242924
Spotted a mistake? Suggest a correction
More in ENT (Otolaryngology)
Radiosensitising strategies in locally advanced head and neck cancer: a systematic review and network meta-analysis of randomised controlled trials (RADIANCE-NMA)
ResearchInt Forum Allergy Rhinol · Oct 18, 2026 · ENT (Otolaryngology)
Changes in Peripheral Immune Cell Subsets Following Endoscopic Sinus Surgery Differ Between CRSwNP and AERD
ResearchJAMA Otolaryngol · Oct 15, 2026 · ENT (Otolaryngology)
Longitudinal Growth of Mandated Hearing Aid Benefits in the US
ResearchJAMA Otolaryngol · Oct 8, 2026 · ENT (Otolaryngology)
Precision Phenotyping With Audiometric Data and Gene Discovery for Sensorineural Hearing Loss
ResearchOtolaryngol Head Neck Surg · Oct 6, 2026 · ENT (Otolaryngology)