Visual abstractsVisuals
ResearchThe Laryngoscope · Jun 4, 2026

Cochlear Implants and Electrocautery: A Live Animal Model

Jeyakumar A, Lu Q, Johnson R, Mitchell R, Brickman T
ENT (Otolaryngology)Basic science

Study type: Laboratory or animal research: results may not apply to people yet.

Read the full paper free at PubMed Central ↗

Abstract

Objectives The use of monopolar cautery is regarded as a contraindication in cochlear implant (CI) recipients, despite a large amount of contradicting data. The objectives are as follows: (1) Evaluate the effects of high-powered electrocautery on CI devices. (2) Record any high-powered electrocautery induced temperature changes. Methods After IACUC approval, 12 manufacturer-donated CIs were implanted in two anesthetized pigs. All devices underwent pre-study integrity checks; six had onsite impedance testing, while six received post-cautery analysis. Fifty watts coagulation was applied for 20 min at sites 1-15 cm from the CI. Bluetooth thermometers monitored three sites of temperatures. Devices were explanted for analysis. Stata 17.0 was used for statistical analysis. A p-value Results The mean temperatures at the three sites (a*, b*, c*) was 32.9°C-38.8°C. Maximum absolute temperature changes averaged 4.1°C (95% CI: 1.4-6.8) for scalp measurements and 2.9°C (95% CI: 1.0-4.9) for round window measurements. Neither time (p = 0.09) nor distance (p = 0.64) generated significant scalp temperature changes. Devices were stratified by proximity. Mean maximum temperature changes showed a gradient by distance. Differences in device proximity were not statistically significant. Postimplantation device analysis was performed on six devices, and all passed the manufacturer in-depth device analysis. Conclusions This controlled experimental evidence, demonstrating 100% device integrity across all tested conditions, challenges current manufacturer contraindications and supports evidence-based guideline revision. Level of evidence N/A.

Topics

Animals; Electrocoagulation; Cochlear Implants; Swine; Models, Animal; Cochlear Implantation

DOI: 10.1002/lary.70649 · PubMed: 42242924

Spotted a mistake? Suggest a correction

More in ENT (Otolaryngology)

Meta-analysisVisual abstracteClinicalMedicine · Oct 26, 2026 · Oncology, ENT (Otolaryngology)

Weekly cisplatin and platinum-based chemoradiotherapy improve overall survival in head and neck cancer

Radiosensitising strategies in locally advanced head and neck cancer: a systematic review and network meta-analysis of randomised controlled trials (RADIANCE-NMA)

ResearchInt Forum Allergy Rhinol · Oct 18, 2026 · ENT (Otolaryngology)

Immune cell changes after sinus surgery differ between two conditions

Changes in Peripheral Immune Cell Subsets Following Endoscopic Sinus Surgery Differ Between CRSwNP and AERD

ResearchJAMA Otolaryngol · Oct 15, 2026 · ENT (Otolaryngology)

Mandated hearing aid coverage rose to 52.5% for US youth and 19.2% for adults

Longitudinal Growth of Mandated Hearing Aid Benefits in the US

ResearchJAMA Otolaryngol · Oct 8, 2026 · ENT (Otolaryngology)

Precision phenotyping identified genetic variants for sensorineural hearing loss better than diagnostic codes

Precision Phenotyping With Audiometric Data and Gene Discovery for Sensorineural Hearing Loss

ResearchOtolaryngol Head Neck Surg · Oct 6, 2026 · ENT (Otolaryngology)

Primary Tumor Location and Lymphatic Spread in Papillary Thyroid Carcinoma