Continuous glucose monitoring did not change glucose control but reduced recurrent low blood sugar in hospitalized patients
Continuous Glucose Monitoring-Guided Insulin Administration in Hospitalized Patients With Diabetes: A Randomized Clinical Trial
Clinical question: Does continuous glucose monitoring improve insulin therapy in hospitalized patients with diabetes?
Why it matters: The efficacy and safety of using continuous glucose monitoring to adjust insulin therapy for hospitalized patients have not been evaluated.
- Researchers randomized 185 general medicine and surgery patients with type 1 and type 2 diabetes to receive insulin adjustments based on either continuous glucose monitoring or standard capillary glucose testing.
- There were no significant differences in time in range (54.51% ± 27.72 vs. 48.64% ± 24.25) or mean daily glucose (183.2 ± 40 vs. 186.8 ± 39 mg/dL) between groups.
- Among patients with hypoglycemia, the continuous glucose monitoring group had fewer recurrent events (1.80 ± 1.54 vs. 2.94 ± 2.76 events/patient) and a lower percentage of time below 70 mg/dL (1.89% ± 3.27 vs. 5.47% ± 8.49).
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Cite this article
Spanakis EK, Urrutia A, Galindo RJ, et al. Continuous Glucose Monitoring-Guided Insulin Administration in Hospitalized Patients With Diabetes: A Randomized Clinical Trial. Diabetes Care. 2022. doi:10.2337/dc22-0716