Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine-SSC/ISTH working group on prothrombin time/international normalized ratio standardization

van den Besselaar AMHP, Stavelin A, Kitchen S, Bryant M, Tripodi A, Scalambrino E, et al
HematologyPathology & Laboratory MedicineNutritionAdult
Read the full paper free at the publisher ↗

Abstract

Calibration of prothrombin time (PT) in terms of international normalized ratio (INR) has been outlined in "Guidelines for thromboplastins and plasmas used to control oral anticoagulant therapy" (World Health Organization, 2013). The international standard ISO 17511:2020 presents requirements for manufacturers of in vitro diagnostic (IVD) medical devices (MDs) for documenting the calibration hierarchy for a measured quantity in human samples using a specified IVD MD. The objective of this article is to define an unequivocal, metrologically traceable calibration hierarchy for the INR measured in plasma as well as in whole blood samples. Calibration of PT and INR for IVD MDs according to World Health Organization guidelines is similar to that in cases where there is a reference measurement procedure that defines the measurand for value assignment as described in ISO 17511:2020. We conclude that, for PT/INR standardization, the optimal calibration hierarchy includes a primary process to prepare an international reference reagent and measurement procedure that defines the measurand by a value assignment protocol conforming to clause 5.3 of ISO 17511:2020. A panel of freshly prepared human plasma samples from healthy adult individuals and patients on vitamin K antagonists is used as a commutable secondary calibrator as described in ISO 17511:2020. A sustainable metrologically traceable calibration hierarchy for INR should be based on an international protocol for value assignment with a single primary reference thromboplastin and the harmonized manual tilt tube technique for clotting time determination. The primary international reference thromboplastin reagent should be used only for calibration of successive batches of the secondary reference thromboplastin reagent.

Topics

Adult; Humans; Prothrombin Time; International Normalized Ratio; Thromboplastin; Calibration; Chemistry, Clinical; Anticoagulants; Reference Standards; Fibrinolytic Agents; Indicators and Reagents; Communication

DOI: 10.1016/j.jtha.2023.12.013 · PubMed: 38128762

Cite this article

van den Besselaar AMHP, Stavelin A, Kitchen S, Bryant M, Tripodi A, Scalambrino E. Defining a metrologically traceable and sustainable calibration hierarchy of international normalized ratio for monitoring of vitamin K antagonist treatment in accordance with International Organization for Standardization (ISO) 17511:2020 standard: communication from the International Federation of Clinical Chemistry and Laboratory Medicine-SSC/ISTH working group on prothrombin time/international normalized ratio standardization. J Thromb Haemost. 2023. doi:10.1016/j.jtha.2023.12.013

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

J Thromb Haemost · Feb 6, 2026 · Research · Hematology

A 25-year evaluation of direct local international normalized ratio calibration of the Owren type prothrombin time method in Sweden

J Thromb Haemost · Mar 27, 2026 · Research · Hematology

Toward a ready-to-use blood collection tube preventing direct oral anticoagulant-interference with routine coagulation assays

Blood Adv · May 27, 2026 · Research · Hematology

A comprehensive multicenter study assessing the impact of abelacimab, asundexian, and milvexian on coagulation assays

Am J Emerg Med · Dec 10, 2023 · Meta-analysis · Emergency Medicine, Nutrition

Safety and efficacy of fixed versus variable-dose prothrombin complex concentrate for emergent reversal of vitamin K antagonists: A systematic review and meta-analysis

Crit Care Med · Feb 14, 2024 · Meta-analysis · Critical Care, Nutrition

Fixed- Versus Variable-Dose Prothrombin Complex Concentrate for the Emergent Reversal of Vitamin K Antagonists: A Systematic Review and Meta-Analysis