Detection of Microvascular Failure Using Bioimpedance and Sensor-derived Clinical Indicators: A Preliminary Study
Study type: Laboratory or animal research: results may not apply to people yet.
Abstract
Background
Early detection of flap failure is critical for successful salvage surgery outcomes. Delays in diagnosis can lead to additional surgical intervention, tissue necrosis, or even flap loss. Current monitoring methods rely heavily on intermittent clinical evaluations, which are subjective and time-consuming and may miss early signs of failure. A continuous, objective monitoring system could significantly improve detection accuracy and response times. This study explores a multisensory strategy to identify early physiological changes associated with flap failure. Methods
Using a rat hindlimb model, flap failure was simulated via femoral vein or artery clamping (n = 5 per group). Physiological responses-including temperature, skin color, and bioimpedance-were recorded continuously for 2 hours postclamping with sensors. Statistical significance was set at a P value less than 0.05 using a 1-sample t test. Distinct failure patterns were compared against nonclamped controls. Results
Clear physiological trends emerged in failure groups. We preliminarily validated a multisensor approach for free flap monitoring in a rat model using temperature, bioimpedance, and color sensing. Each modality provided distinct and complementary insights into perfusion changes: color sensing detected changes fastest but was more variable, temperature changes were moderate and potentially influenced by external heating, and bioimpedance showed the most consistent trends over time, highlighting the value of a combined sensing strategy for reliable tissue viability assessment. Conclusions
This multisensory approach demonstrates strong potential for early, noninvasive flap failure detection through continuous monitoring of temperature, color, and bioimpedance. These findings support further development of real-time sensing systems to enhance surgical outcomes and patient safety in flap-based procedures.
Spotted a mistake? Suggest a correction
More in Plastic & Reconstructive Surgery
Releasable plating systems had higher complication rates than continuous systems
Continuous Versus Releasable Plating Systems for Sternal Closure: A Systematic Review of Outcomes
Saphenous nerve terminated near navicular cuneiform joint in 45% of specimens
Evaluating Saphenous Nerve Variations With Classification System in the Lower Limb, an Anatomical Study With Surgical Relevance
Evaluation of an Animated Patient Information Video: Improving Consent and Decision-making for Deep Inferior Epigastric Perforator Flap Breast Reconstruction
Most plastic surgery residents satisfied with elective rotation experience
Optimizing the Chief Resident Elective Rotation: Insights From Alumni Feedback in Plastic Surgery Training