Development and Analytical Validation of a Multiplex LC-MRM-MS-Based Reference Measurement Procedure for Apolipoprotein A-I and Btotal Quantification in Serum
Diederiks NM, Romijn FPHTM, van Neer NJM, Pieterse M, Dittrich J, Ceglarek U, et al
Abstract
Background
Medical results from regulated in vitro diagnostic devices should be accurate. The previous reference measurement procedures (RMP) for apolipoprotein A-I (ApoA-I)/apolipoprotein B total (ApoBtotal) standardization are no longer functional. A new RMP aligned with ISO15193:2009 and calibration models of ISO17511:2020 was developed using multiplex liquid chromatography-multiple reaction monitoring-mass spectrometry (LC-MRM-MS) for quantifying ApoA-I and ApoBtotal in serum. Methods
Serum samples underwent reduction, alkylation, lysyl endopeptidase/trypsin digestion, and proteotypic peptide purification. Optimal retention times, precursor/fragment ion m/z were selected using an Agilent-1290 ultra-high-performance liquid chromatography coupled to a 6495C triple-quadrupole mass spectrometer. Calibrators were indirectly traceable to WHO/IFCC SP1-01/SP3-08 (ApoA-I/ApoBtotal) via the RMP, with propagated measurement uncertainty (MU), supported by replicate SP1-01/SP3-08 measurements across independent batches. Validation followed ISO15193:2009 and included linearity, recovery, uncertainty, precision, stability, preanalytical conditions, carryover, limit of blank (LoB), limit of detection (LoD), limit of quantification (LoQ), and comparability. Results
Linearity was confirmed from 0.10 to 2.97 g/L and 0.27 to 3.83 g/L (ApoA-I/ApoBtotal), with Conclusions
This LC-MRM-MS-based RMP for ApoA-I and ApoBtotal provides a next-generation RMP succeeding the former ELISA-based RMP. Certification of commercially available ApoA-I/ApoBtotal tests will enable continued global standardization of ApoA-I/ApoBtotal measurements, while preserving traceability of test results to the former WHO-IFCC reference materials within maximum allowable uncertainty.
Topics
Humans; Apolipoprotein A-I; Liquid Chromatography-Mass Spectrometry; Reference Standards; Apolipoproteins B; Calibration; Chromatography, High Pressure Liquid; Chromatography, Liquid; Reproducibility of Results; Mass Spectrometry
DOI: 10.1093/clinchem/hvag041 · PubMed: 42201254
Spotted a mistake? Suggest a correction
More in Pathology & Laboratory Medicine
ResearchMod Pathol · Nov 18, 2026 · Pathology & Laboratory Medicine
Intra-Patient Cell-Surface Receptor Normalization Refines TNBC Biomarker Profiling for Precision Oncology
ResearchMod Pathol · Nov 18, 2026 · Pathology & Laboratory Medicine
CD40 expression is highly sensitive and specific for differentiating between classic Hodgkin lymphoma and ALK-negative anaplastic large cell lymphoma
ResearchMod Pathol · Nov 12, 2026 · Pathology & Laboratory Medicine
ResearchPediatricMod Pathol · Nov 12, 2026 · Pathology & Laboratory Medicine
ResearchMod Pathol · Nov 12, 2026 · Pathology & Laboratory Medicine
Pilomatrix-like High-Grade Endometrioid Carcinoma in the Female Genital Tract: A Single-Institution Clinicopathologic Analysis of 15 Cases