ReviewResuscitation · Jul 1, 2026

Device damage and malfunction during dual defibrillation or dual electrical cardioversion: a scoping review

Purkarthofer D, Bachner V, Labenbacher S, Gruber J, Silwanis C, Scquizzato T, et al
Critical CareAdult
Read the full paper free at the publisher ↗

Abstract

Purpose To map and characterize device damage and malfunction associated with dual defibrillator use in defibrillation and cardioversion, which is increasingly employed in both elective and emergency settings, and to identify knowledge gaps relevant to practice and future research. Methods We conducted a scoping review in accordance with JBI methodology and PRISMA-ScR. PubMed, Embase, CINAHL, CENTRAL, and gray literature were searched without language or date restrictions. Regulatory adverse event databases were also searched. Eligible sources reported device damage or malfunction associated with two external defibrillators delivering shocks within one second or less of each other. Results Twenty-nine records from 2016 to 2026 were included: five published sources and 24 regulatory adverse event reports. Published sources comprised two randomized trials, one survey study, one case report, and one ad-hoc retrospective study. Across five published sources, eight events of lasting device damage and four temporary malfunctions were identified. In published studies, all reported damage events were reported in association with double simultaneous defibrillation or dual cardioversion. Across 24 regulatory reports, 11 records described device damage and 16 described malfunctions. Reported damage was often not recognized during the index event and was instead identified on post-use testing or later routine self-checks. Discussion Device damage and malfunction during dual defibrillator use are documented in both published and regulatory sources. Damage may remain occult until later testing, creating risk beyond the index case. Standardized prospective reporting and routine post-use device testing should be considered wherever dual defibrillator use is employed.

Topics

Humans; Electric Countershock; Defibrillators; Equipment Failure

DOI: 10.1016/j.resuscitation.2026.111186 · PubMed: 42385983

Spotted a mistake? Suggest a correction

More in Critical Care

Meta-analysisCrit Care · Nov 19, 2026 · Critical Care

Corticosteroids in sepsis have conflicting mortality conclusions in reviews

Discordant evidence on corticosteroids in sepsis: a meta-research study

ResearchPediatricResuscitation · Nov 17, 2026 · Critical Care

Eighty-nine percent of infants survived 28 days after chest compressions at birth in 2024

Survival after neonatal chest compressions at birth: Temporal trends and prognostic factors in a population-based cohort, Queensland, Australia

ResearchResuscitation · Nov 17, 2026 · Critical Care

Bystanders attaching AEDs had highest 30-day survival rate of 28 percent

Survival after out of hospital cardiac arrest in relation to AED usage in a multi response system including AED equipped drones

ResearchResuscitation · Nov 17, 2026 · Critical Care

Foreign body airway obstruction caused 1.1% of out-of-hospital cardiac arrests

Out-of-hospital cardiac arrests due to foreign body airway obstruction: A CanROC registry, retrospective cohort study

ResearchResuscitation · Nov 17, 2026 · Critical Care

Commercially available multimodal LLMs for EEG-based prognostication after cardiac arrest