ReviewResuscitation · Jul 1, 2026
Device damage and malfunction during dual defibrillation or dual electrical cardioversion: a scoping review
Purkarthofer D, Bachner V, Labenbacher S, Gruber J, Silwanis C, Scquizzato T, et al
Abstract
Purpose
To map and characterize device damage and malfunction associated with dual defibrillator use in defibrillation and cardioversion, which is increasingly employed in both elective and emergency settings, and to identify knowledge gaps relevant to practice and future research. Methods
We conducted a scoping review in accordance with JBI methodology and PRISMA-ScR. PubMed, Embase, CINAHL, CENTRAL, and gray literature were searched without language or date restrictions. Regulatory adverse event databases were also searched. Eligible sources reported device damage or malfunction associated with two external defibrillators delivering shocks within one second or less of each other. Results
Twenty-nine records from 2016 to 2026 were included: five published sources and 24 regulatory adverse event reports. Published sources comprised two randomized trials, one survey study, one case report, and one ad-hoc retrospective study. Across five published sources, eight events of lasting device damage and four temporary malfunctions were identified. In published studies, all reported damage events were reported in association with double simultaneous defibrillation or dual cardioversion. Across 24 regulatory reports, 11 records described device damage and 16 described malfunctions. Reported damage was often not recognized during the index event and was instead identified on post-use testing or later routine self-checks. Discussion
Device damage and malfunction during dual defibrillator use are documented in both published and regulatory sources. Damage may remain occult until later testing, creating risk beyond the index case. Standardized prospective reporting and routine post-use device testing should be considered wherever dual defibrillator use is employed.
Topics
Humans; Electric Countershock; Defibrillators; Equipment Failure
DOI: 10.1016/j.resuscitation.2026.111186 · PubMed: 42385983
Spotted a mistake? Suggest a correction
More in Critical Care
Meta-analysisCrit Care · Nov 19, 2026 · Critical Care
Discordant evidence on corticosteroids in sepsis: a meta-research study
ResearchPediatricResuscitation · Nov 17, 2026 · Critical Care
Survival after neonatal chest compressions at birth: Temporal trends and prognostic factors in a population-based cohort, Queensland, Australia
ResearchResuscitation · Nov 17, 2026 · Critical Care
Survival after out of hospital cardiac arrest in relation to AED usage in a multi response system including AED equipped drones
ResearchResuscitation · Nov 17, 2026 · Critical Care
Out-of-hospital cardiac arrests due to foreign body airway obstruction: A CanROC registry, retrospective cohort study
ResearchResuscitation · Nov 17, 2026 · Critical Care