Low-cost cardboard incubator maintained infant body temperature as well as standard incubators
Disposable low-cost cardboard incubator for thermoregulation of stable preterm infant - a randomized controlled non-inferiority trial
Open-labeled non-inferiority randomized controlled trialBenefit found
Can a low-cost cardboard incubator maintain thermoregulation in stable preterm infants?
Who
96
Stable preterm infants at 32 to 36 weeks post-menstrual age · Tertiary level teaching hospital in India
Compared
Standard single-wall incubator: 0% failed thermoregulation
Low-cost cardboard incubator: 0% failed thermoregulation
Main outcome
Maintenance of skin and axillary temperatures
Result: -0.07 °C (95% CI -0.11 to -0.04) for skin and -0.06 °C (95% CI -0.095 to -0.02) for axillary
Bottom line: A low-cost cardboard incubator is non-inferior to standard single-wall incubators for maintaining thermoregulation in stable preterm infants.
eClinicalMedicine · 2020-12-07 · Funded by Food and Drug Administration
Why it matters: Standard incubators costing 1,500 to 35,000 USD are often unaffordable for many medical units in low and middle-income countries.
Headline and visual abstract written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Cite this article
Chandrasekaran A, Amboiram P, Balakrishnan U, et al. Disposable low-cost cardboard incubator for thermoregulation of stable preterm infant - a randomized controlled non-inferiority trial. EClinicalMedicine. 2021;31:100664. doi:10.1016/j.eclinm.2020.100664
Spotted a mistake? Suggest a correction