No tested app-linked feedback device provided overall adequate cardiopulmonary resuscitation performance for laypersons
Do laypersons need App-linked real-time feedback devices for effective resuscitation? - Results of a prospective, randomised simulation trial
Why it matters: It is unknown how different app-linked real-time feedback devices currently on the market compare to each other regarding the quality of resuscitation performed by laypersons.
- Forty participants performed compression-only cardiopulmonary resuscitation on a manikin using four different app-linked feedback devices compared to a baseline without feedback.
- SimCPR®ProTrainer was the only device associated with guideline-compliant chest compressions and improved chest compression depth (5.37 ± 0.76) and percentage of correct depth (p < 0.001).
- CorPatch® Trainer was the only device associated with improved chest recoil (72.25 ± 24.89) compared to baseline (49.00 ± 42.20), while other devices were associated with lower recoil rates.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Cite this article
Wingen S, Großfeld N, Adams NB, Streit A, Stock J, Böttiger BW. Do laypersons need App-linked real-time feedback devices for effective resuscitation? - Results of a prospective, randomised simulation trial. Resusc Plus. 2024. doi:10.1016/j.resplu.2024.100631