Early Clinical Experience With Robot-Assisted Electrode Insertion in Cochlear Implantation Using the Otoarm-Otodrive System

David KJV, Lin VYW, Chen JM, Spiegel JL, Ng AH, Smilsky KL, et al
ENT (Otolaryngology)Adult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Importance Manual cochlear implant (CI) electrode insertion remains surgeon-dependent and may contribute to intracochlear trauma. Robotic-assisted systems, such as the new Otoarm-Otodrive, offer motorized control at ultra-slow speeds that aim to minimize force fluctuations and pressure changes, potentially improving hearing preservation and surgical consistency. Objective To describe the institutional experience, surgical workflow integration, and feasibility of robot-assisted cochlear implantation (CI) using the Otoarm-Otodrive system. Design Retrospective descriptive study of 50 consecutive robot-assisted CI cases. Setting Tertiary referral otology center in Canada. Participants Fifty Adult patients undergoing cochlear implantation without cochlear malformation or ossification using a MED-EL device from January to October 2025. Intervention Robot-assisted cochlear electrode array insertion using the Otoarm-Otodrive (MED-EL) system. Main outcome measures Feasibility (setup and insertion times), insertion depth, intraoperative challenges, and integration into operative workflow. Results The additional time for robot setup and alignment averaged 374 seconds. Mean electrode insertion duration, performed at a standardized rate of 0.1 mm/second, was 444 seconds. Postoperative Otoplan analysis in 30 cases demonstrated a mean angular insertion depth of 575°. No major complications such as facial nerve palsy or infections, and no device-related complications occurred. All but 1 case achieved full insertion. Conclusions Robot-assisted cochlear electrode insertion using the Otoarm-Otodrive system is safe, reproducible, and easily integrated into standard CI workflow. Relevance This first North American experience in using the Otoarm-Otodrive system demonstrates the feasibility of robotic cochlear implant insertion as a step toward achieving atraumatic implantation. Future studies will assess hearing preservation benefits and cost-effectiveness for broader adoption.

Topics

Humans; Cochlear Implantation; Retrospective Studies; Robotic Surgical Procedures; Female; Adult; Feasibility Studies; Male; Middle Aged; Cochlear Implants; Aged; Electrodes, Implanted

DOI: 10.1177/19160216261460457 · PubMed: 42708541

Cite this article

David KJV, Lin VYW, Chen JM, Spiegel JL, Ng AH, Smilsky KL. Early Clinical Experience With Robot-Assisted Electrode Insertion in Cochlear Implantation Using the Otoarm-Otodrive System. J Otolaryngol Head Neck Surg. 2026. doi:10.1177/19160216261460457

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Laryngoscope · Jan 21, 2026 · Research · ENT (Otolaryngology)

Robotic-Assisted Electrode Array Insertion Improves Stability of Acoustic Hearing Thresholds

Laryngoscope · Apr 14, 2026 · Research · ENT (Otolaryngology)

Micro-CT Evaluation of Robot-Assisted Insertion of CI632 Implant in Human Cadaveric Temporal Bones

J Otolaryngol Head Neck Surg · May 6, 2026 · Research · ENT (Otolaryngology), Surgery

Evaluating the Safety and Efficacy of Robot-Assisted Cochlear Implant Electrode Array Insertion

J Neuroeng Rehabil · Jun 30, 2026 · Review · Physical Medicine & Rehabilitation, ENT (Otolaryngology)

Biohybrid cochlear implants: neural interfaces, regenerative pathways, and translational benchmarks

Otolaryngol Head Neck Surg · Jun 17, 2026 · Meta-analysis · ENT (Otolaryngology), Surgery Visual abstract

Cochlear implant complication rates differ between lower-middle and upper-middle income countries