Visual abstractsVisuals
ResearchJournal of the American Heart Association · Dec 17, 2025

ECG-Capable Smartwatches Can Induce Magnet Mode in Cardiac Implantable Electronic Devices

Wegner FK, Geisendörfer T, Wolfes J, Rath B, Köbe J, Ellermann C, et al
CardiologyBasic science

Study type: Laboratory or animal research: results may not apply to people yet.

Read the full paper free at PubMed Central ↗

Abstract

Background Smartwatches are increasingly used for screening of tachyarrhythmias and for ECG recording. We therefore investigated possible mechanisms of magnet-mode induction of smartwatches in cardiac implantable electronic devices (CIED). Methods Fifteen CIED (5 pacemakers, 10 implantable cardioverter-defibrillators) from all manufacturers were implanted in a subcutaneous and submuscular location in an isolated porcine thorax and connected to an interactive heart simulator. Eight different smartwatches (3 Apple, 3 Samsung, 1 Huawei, 1 Withings) were placed on top of the implantation site in 3 different configurations, and signs of magnet mode induction were recorded. Results When the faces of smartwatches were placed on top of subcutaneously implanted CIED, only 1 instance of magnet mode induction was recorded in 120 measurements (<1%). When turning around the smartwatches and placing the backs on the implantation site, signs of magnet mode were recorded in 36 of 120 measured instances (30%). Smartwatches connected to respective chargers induced magnet mode in 28 of 120 measurements (23%). Magnet mode induction in submuscularly implanted CIED was seen in 5 of 360 measurements (1%). The smartwatch with physical charging connectors was the only one not to induce magnet mode in any CIED, resulting from a lower-than-average magnetic field strength likely due to its different charging mechanism. Conclusions ECG-capable smartwatches can induce magnet mode in CIED. Although the risk for patients wearing smartwatches on their wrist is low, patients should be discouraged from placing their watch close to their CIED. Smartwatch charging mechanism and CIED implantation depth influence the risk of magnet mode induction.

Topics

Defibrillators, Implantable; Pacemaker, Artificial; Animals; Electrocardiography; Swine; Equipment Design; Magnets

DOI: 10.1161/JAHA.125.044972 · PubMed: 41404751

Cite this article

Wegner FK, Geisendörfer T, Wolfes J, Rath B, Köbe J, Ellermann C. ECG-Capable Smartwatches Can Induce Magnet Mode in Cardiac Implantable Electronic Devices. J Am Heart Assoc. 2025. doi:10.1161/JAHA.125.044972

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Eur J Cardiothorac Surg · Sep 16, 2026 · Research · Cardiothoracic & Vascular Surgery

TricValve System Implantation in Patients With Cardiac Implantable Electronic Devices: A Multicentre Observational Study

J Am Heart Assoc · Mar 13, 2026 · Research · Cardiology

Infective Endocarditis After Arrythmia Device Implantation in Cardiac Surgery Patients

J Am Heart Assoc · Jun 23, 2026 · Research · Cardiology

Mental Disorders After Cardiac Implantable Electronic Device Implantation in Young Individuals

J Am Heart Assoc · Aug 20, 2026 · Research · Cardiology

Individual and Community-Level Differences in Use of Contemporary Electrophysiology Procedures in the United States

J Med Internet Res · Nov 13, 2026 · Meta-analysis · Digital Health & AI Visual abstract

Remote monitoring of heart devices shows potential economic value in some settings