Objective
To assess intracochlear electrocochleography (ECochG) for estimating cochlear implant (CI) performance for lateral wall and perimodiolar electrode arrays. Study design
Prospective cohort study. Setting
Tertiary referral center. Methods
From December 2021 to June 2024, ECochG was measured intraoperatively using the Active Insertion Monitoring (AIM) system by Advanced Bionics. After insertion, multifrequency tone bursts were delivered via a sound tube from 250 Hz to 2 kHz at 95 to 105 dB HL with recordings at each internal electrode. Using spectral analysis, the maximal amplitude (MA) response electrode was identified by frequency and summed into a single value of total response (MA-ECochG-TR). Postoperative CI performance was routinely monitored. Results
Of 57 participants with ECochG recordings, 43 completed 6-month testing. Of these, 23 were perimodiolar (Mid-Scala [MS]), and 20 were lateral wall (SlimJ [SJ]). MA-ECochG-TR had a moderately strong positive correlation to 6-month consonant-nucleus-consonant (CNC) word scores for MS cases (r = 0.73, 95% CI [0.46-0.88], P Conclusion
MA-ECochG-TR independently estimates performance for perimodiolar arrays. Intracochlear ECochG is less predictive for lateral wall arrays; this may be due to confounding from EAS (though this was not statistically significant) or from signal degradation with a larger electrode-modiolar distance.