GAD65 Antibody ELISA With Extended Reportable Range: Validation and Guidance for Neurological Practice
Abstract
Objective
To (1) validate GAD65-ELISA detection and quantification for type 1 diabetes mellitus and autoimmune neurological diagnoses, (2) correlate ELISA results (reference range Methods
Serum performance for diabetes (FDA-cleared, undiluted) was verified, and neurological laboratory-developed serum and CSF dilution protocols were validated to extend the reportable range beyond > 250 IU/mL. ELISA and RIA values were correlated, including established neurological RIA cut-offs (serum ≥ 20 nmol/L; CSF any positive). Results
ELISA met analytical criteria (precision, accuracy, sensitivity, specificity, reference range) in serum and CSF and was clinically equivalent to RIA for autoimmune diabetes. Neurological ELISA cut-offs were established at 10,000 IU/mL (serum) and 100 IU/mL (CSF). Precision was better below 10,000 IU/mL (CVs 10,000 IU/mL. Clinical specificity was 97.5% in serum and 100% in CSF, exceeding reported RIA specificity. Screen results > 250 IU/mL spanned a wide range of dilution values; many were below the neurological cut-off. Most patients with paired serum/CSF positivity showed elevated GAD65 IgG indices. Interpretation
GAD65 ELISA and RIA have equivalent sensitivity for autoimmune diabetes and neurological testing, with higher specificity for ELISA. A serum cut-off of 10,000 IU/mL is informative but requires clinical context, and dilution of screen > 250 IU/mL samples is essential for neurological interpretation.