Meta-analysisJournal of Pediatrics · Dec 13, 2024

Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia

St Clair SL, Ulyatt CM, Corkin MT, Lord LG, Crowther CA, Harding JE, et al
PediatricsEndocrinology & DiabetesPediatric
Read the full paper free at the publisher ↗

Abstract

Objective To evaluate the accuracy of various point-of-care device methodologies for measuring blood glucose concentrations in babies at risk of neonatal hypoglycemia. Study design This systematic review and meta-analysis included studies from Ovid MEDLINE, Ovid Embase, and Web of Science up to May 20, 2024. Studies comparing point-of-care testing methods for neonatal blood glucose to a standard laboratory method were included, excluding those on continuous glucose monitoring or conducted before 1990. Two researchers independently assessed inclusion and evaluated risk of bias using QUADAS-2. Sensitivity and specificity were calculated using contingency tables, and diagnostic accuracy was analyzed using hierarchical random-effects modelling. Studies with insufficient data were summarized by estimation direction. Results Seventy-one studies were included. The quantitative analysis (n = 31) evaluated glucose oxidase (GO) + photometry (n = 8), glucose-1-dehydrogenase (GDH) + photometry (n = 6), GO + electrochemistry (n = 13), GDH + electrochemistry (n = 12), and hexokinase (HK) + electrochemistry (n = 2). All methods showed high specificity (≥93%), with GO + electrochemistry, GDH + electrochemistry, and HK + electrochemistry showing superior sensitivity. The summary receiver operating characteristic curve confirmed HK + electrochemistry as the most accurate. Conclusion Certain point-of-care device methodologies demonstrate greater accuracy in measuring neonatal blood glucose concentrations. Of the methods evaluated, HK + electrochemistry proved to be the most reliable. However, the limited number of studies using this method suggests the need for further research to confirm these findings across diverse settings and populations. Systematic review registration PROSPERO on December 29, 2023 (CRD42023488539).

Topics

Humans; Infant, Newborn; Hypoglycemia; Blood Glucose; Point-of-Care Testing; Sensitivity and Specificity; Point-of-Care Systems; Infant, Newborn, Diseases

DOI: 10.1016/j.jpeds.2024.114438 · PubMed: 39675663 · Registered: PROSPERO CRD42023488539

Cite this article

St Clair SL, Ulyatt CM, Corkin MT, Lord LG, Crowther CA, Harding JE. Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia. J Pediatr. 2024. doi:10.1016/j.jpeds.2024.114438

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

JAMA · Dec 12, 2023 · Randomized trial · Endocrinology & Diabetes Visual abstract

Adding metformin to insulin did not change composite neonatal outcomes in pregnant people with diabetes

JAMA Pediatr · Oct 2, 2023 · Meta-analysis · Pediatric · Pediatrics, Endocrinology & Diabetes

Infants Eligible for Neonatal Hypoglycemia Screening: A Systematic Review

Hepatol Commun · Apr 24, 2026 · Research · Pediatric · Gastroenterology & Hepatology, Pediatrics

Differentiation of biliary atresia using a rapid point-of-care matrix metalloproteinase-7 assay: A multicenter diagnostic study

Diabetologia · Sep 22, 2026 · Guideline · Endocrinology & Diabetes

International experts propose practical guidelines for using diabetes technology in non-critically ill hospitalized adults

Diabetes Care · Sep 3, 2025 · Randomized trial · Endocrinology & Diabetes Visual abstract

Continuous glucose monitoring effects on hypoglycemia awareness vary depending on the scoring method used