Impact of Device Insertion Technique on Impella CP Position and Orientation Within the Left Ventricle

Baldetti L, Beneduce A, Vanwersch C, Sonntag SJ, Ferreira de Almeida D, Knabe M, et al
CardiologyAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Background Impella CP malrotation diminishes the unloading effect of the device. There is a paucity of data informing on the insertion procedural factors associated with malrotation. Objectives The aim of the study was to analyze the impact of each device insertion step on the final orientation of the Impella CP within the left ventricle using an ex vivo simulator. Methods Two expert operators ran a prespecified set of Impella CP insertion procedures (n = 96) across different anatomical silicon models printed from electrocardiogram-gated computed tomography scans obtained during Impella support (n = 6). One operator ran Impella CP insertion blinded to the simulator, while the other assessed the position and orientation of the device in real time under direct visualization and was considered the gold standard. The operators then swapped their roles. Results All 96 preplanned insertions were successfully carried out. Impella malrotation rate was 45.8%. The lowest rate of malrotation was found when the device was inserted in the femoral sheath with its convex elbow at 9 o'clock (16.7%; P = 0.008). At the end of insertion procedure, a "deep" position was observed more frequently in optimally oriented cases (98.1% vs 70.5%; P < 0.001). Finally, a pull-back maneuver to bring the radiopaque marker at the level of the aortic valve, as recommended, led to malrotation in 99.0%. Conclusions In this preclinical study malrotation occurred in 45.8%, with no interaction with operator or anatomical model. Pump orientation in the introducer sheath and insertion depth in the left ventricle were associated with malrotation outcome. This experiment identified actionable targets to prevent Impella CP malrotation that warrant validation in clinical studies.

DOI: 10.1016/j.jacadv.2025.102581 · PubMed: 41616588

Cite this article

Baldetti L, Beneduce A, Vanwersch C, Sonntag SJ, Ferreira de Almeida D, Knabe M. Impact of Device Insertion Technique on Impella CP Position and Orientation Within the Left Ventricle. JACC Adv. 2026. doi:10.1016/j.jacadv.2025.102581

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

More in Cardiology

JACC Adv · Nov 23, 2026 · Research · Cardiology

Building a Rural Cardiovascular Workforce Through Research Fellowship Immersion: Thad Waites Rural CVD Research Fellowship

Eur Respir J · Nov 23, 2026 · Review · Pulmonology, Cardiology

BMP-9 and BMP-10 regulate pulmonary vascular resilience through complex pathways

JACC Adv · Nov 23, 2026 · Research · Cardiology

NT-proBNP and imaging improve heart failure prediction in European population

JACC Adv · Nov 23, 2026 · Research · Cardiology

Slower walkers had higher risk of atherosclerotic cardiovascular disease than brisk walkers

Ann Fam Med · Nov 23, 2026 · Research · Family Medicine & Primary Care, Endocrinology & Diabetes

Lower Estimated Insulin Sensitivity Linked to Greater Cardiovascular Disease Risk