In Silico Modeling Validation and Contact Pressure Distribution Comparison Analysis of Conventional and Robotic-Assisted Unicompartmental Knee Arthroplasty
Study type: Laboratory or animal research: results may not apply to people yet.
Abstract
Background
Unicompartmental knee arthroplasty (UKA) improves recovery, but optimizing load remains challenging. This study measures tibiofemoral contact pressures after conventional and robotic UKA in cadavers and validates a finite element (FE) model against experimental data. Methods
Sixteen fresh-frozen cadaveric lower limbs underwent medial UKA with conventional cutting guides (n = 8) or CORI robotic assistance (n = 8). Medial (implanted) and lateral (native) compartment pressures were recorded from 0° to 90° using pressure film sensors. The FE model replicated and was validated against these measurements. Results
Medial compartment pressure decreased significantly from 1.864 MPa at 0° to 0.252 MPa at 90° (p 0.05). The FE model demonstrated strong agreement with the measured data, with r2 values of 0.9994 (medial) and 0.9962 (lateral). Conclusion
Conventional and robotic-assisted UKA techniques demonstrated similar contact pressure profiles. However, increased force and area in the native lateral compartment may predispose to postoperative degeneration. The FE model reliably predicted contact behavior and may be especially useful in refining conventional UKA techniques. Level of evidence
Therapeutic Level II. See Instructions for Authors for a complete description of levels of evidence.