Long-shaft vitrectomy probes were associated with lower trocar removal rates in highly myopic eyes
Long-Shaft Vitrectomy Probe for Vitreoretinal Diseases in Highly Myopic Eyes: A Randomized Controlled Trial
Clinical question: Is a long-shaft vitrectomy probe safe and effective in highly myopic eyes?
Why it matters: Standard vitrectomy probes may face difficulty accessing the posterior pole in highly myopic eyes with long axial lengths.
- Researchers conducted a randomized controlled trial of 86 patients with highly myopic eyes, comparing a 31.75 mm long-shaft probe to a 27 mm standard probe.
- The long-shaft probe group had a lower trocar removal rate of 5% compared to 67% in the control group, though instrument bending was higher at 36% versus 14%.
- In eyes with an axial length over 28 mm, the long-shaft probe was associated with better access for removing vitreous gel and trimming the internal limiting membrane.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Cite this article
Hsia Y, Lee CY, Tsui MC, Wang SW, Huang CJ, Ma IH. Long-Shaft Vitrectomy Probe for Vitreoretinal Diseases in Highly Myopic Eyes: A Randomized Controlled Trial. Ophthalmol Sci. 2025. doi:10.1016/j.xops.2025.100824