Five-year safety data shows iStent inject placement is comparable to cataract surgery alone
Long-Term Endothelial Safety Profile With iStent Inject in Patients With Open-Angle Glaucoma
Clinical question: Is iStent inject safe for endothelial cells in primary open-angle glaucoma patients?
Why it matters: It was unknown if the long-term safety and stability of the iStent inject device would affect the density of the corneal endothelial cells.
- The study followed 227 patients with mild-to-moderate primary open-angle glaucoma for 60 months after receiving either iStent inject with cataract surgery or cataract surgery alone.
- At 60 months, the mean percentage decrease in endothelial cell density was 14.3% ± 13.4% in the iStent inject group and 14.8% ± 10.3% in the control group.
- No specific device-related adverse events or complications were reported through month 60.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Cite this article
Ahmed IIK, Sheybani A, De Francesco T, Lass JH, Benetz BA, Samuelson TW. Long-Term Endothelial Safety Profile With iStent Inject in Patients With Open-Angle Glaucoma. Am J Ophthalmol. 2023. doi:10.1016/j.ajo.2023.02.014