Macular Layer Thickness with Spectral-Domain and High-Resolution OCT
Baek SU, Sharpe GP, Shuba LM, Nicolela MT, Chauhan BC
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Objective
The new generation high-resolution OCT (HR-OCT) offers higher axial resolution compared to standard spectral-domain OCT (SD-OCT). This study aimed to determine whether potential differences in the segmentation of retinal layers in the macula with the 2 techniques could lead to clinically meaningful differences with implications for monitoring progression when patients are switched to HR-OCT. Design
Prospective cross-sectional study. Subjects
This study included glaucoma patients and healthy control subjects. Methods
Each eye underwent SD-OCT followed by HR-OCT, with both macular scan patterns acquired at the same transverse location. Automated segmentation was applied and the thickness of 6 individual retinal layers calculated for 1024 A-scans for each of 97 B-scans. Intradevice difference was also assessed with repeated scans. Main outcome measures
Differences in retinal layer thickness measurements between SD-OCT and HR-OCT. Results
A total of 68 eyes (both eyes of 17 glaucoma patients and 17 healthy control subjects) underwent imaging. Mean differences in retinal layer thickness between SD-OCT and HR-OCT ranged from -2.65 to 2.02 μm. Hyper-reflective layers were thicker with SD-OCT, while hyporeflective layers were thicker with HR-OCT, showing an alternating pattern in interdevice differences in thickness measurements. Intradevice differences in thickness measurements, evaluated in a subset of 10 eyes of 5 participants, were within ± 1 μm. Image quality was not an important predictor of the interdevice thickness differences (R2 Conclusions
Interdevice differences in retinal layer thickness were small but consistent. The alternating pattern is likely due to the lower axial resolution of SD-OCT making hyper-reflective layer borders blurred and consequently thicker. For follow-up, patients can be switched to HR-OCT from SD-OCT; however, depending on the magnitude of change required for detection, the differences between the 2 techniques could potentially mask or masquerade change. Financial disclosures
Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
DOI: 10.1016/j.xops.2026.101251 · PubMed: 42437117
Spotted a mistake? Suggest a correction
More in Ophthalmology
ResearchPediatricAm J Ophthalmol · Nov 20, 2026 · Ophthalmology
Association of allergic conjunctivitis severity with meibomian gland abnormalities in children: a cross-sectional study
Low-Concentration Atropine Eye Drops in Children With Myopia and Intermittent Exotropia: The AMIXT2 Randomized Clinical Trial
ResearchAm J Ophthalmol · Nov 19, 2026 · Ophthalmology
Smoke, Not Smoking, as a Possible Risk Factor Triggering Conversion in Leber's Hereditary Optic Neuropathy: A Retrospective Case Series
ResearchOphthalmology · Nov 18, 2026 · Ophthalmology
Trends and Variation in Uveal Melanoma Management in the National Cancer Database, 2004-2023
ResearchAm J Ophthalmol · Nov 18, 2026 · Ophthalmology
Developing Static Perimetry Thresholds to Assess Visual Field Loss Severity in Patients with Retinitis Pigmentosa