Measurement agreement between a portable and a conventional impulse oscillometry system: a two-centre method-comparison study
Yao S, Zhang R, Chen B, Huang L, Lin K, Ji L, et al
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Background
Impulse oscillometry (IOS) enables tidal-breathing assessment of airway mechanics and is useful in patients with limited cooperation. With the growing use of portable IOS devices for bedside and community pulmonary function assessment, agreement with established laboratory systems requires evaluation. Objective
To assess measurement agreement and clinical classification agreement between the portable IOS 600 and the conventional Jaeger MasterScreen-IOS. Methods
In this prospective two-centre method-comparison study with randomised measurement-sequence allocation, adult and paediatric participants underwent IOS measurement using both devices during a single visit. Core parameters (respiratory impedance at 5 Hz (Z5), resistance at 5 Hz (R5), resistance at 20 Hz (R20), reactance at 5 Hz (X5), reactance area and resonant frequency (Fres)) and secondary parameters were analysed. Interdevice agreement was assessed using intraclass correlation coefficients (ICCs), Bland-Altman analysis and coefficients of variation. Clinical classification agreement for core parameters was evaluated using Gwet's agreement coefficient 1 (AC1). Results
Of 200 enrolled participants, 164 were included in the final paired analysis. Core IOS parameters showed high interdevice agreement, with ICCs ranging from 0.902 to 0.961. Bland-Altman analysis demonstrated small mean interdevice differences; however, proportional bias was observed for Z5, R5 and Fres, indicating that agreement at markedly abnormal values should be interpreted cautiously. Coefficients of variation did not differ significantly between devices. Clinical classification agreement was strong to almost perfect, with Gwet's AC1 values of 0.90, 0.92, 0.91 and 0.96 for Z5, R5, R20 and X5, respectively. Adult-only and exploratory disease-subgroup analyses showed generally consistent agreement estimates. Most secondary resistance parameters showed strong-to-almost-perfect agreement, whereas reactance at 15 Hz and reactance at 25 Hz showed lower agreement. Conclusions
The portable IOS 600 demonstrated high measurement agreement and clinical classification agreement with the MasterScreen-IOS for core IOS parameters under standardised testing conditions. Because the final cohort was predominantly adult, these findings are primarily applicable to adult participants. IOS 600 may serve as a complementary tool for airway function assessment, although further validation is needed in paediatric, disease-specific, bedside, primary care and community settings.
Abstract reproduced under the article’s CC BY NC licence.
Topics
Humans; Oscillometry; Prospective Studies; Respiratory Function Tests; Female; Adult; Male; Reproducibility of Results; Child; Middle Aged; Adolescent
DOI: 10.1136/bmjresp-2026-004292 · PubMed: 42409413
Spotted a mistake? Suggest a correction
More in Pulmonology
ReviewEur Respir J · Nov 23, 2026 · Pulmonology
Solving an enigma: the BMP-9/10 axis as a key regulator of pulmonary vascular resilience
Needle aspiration versus intercostal chest drain for secondary spontaneous pneumothorax (the PRINCE-SSP trial): a pragmatic, open-label, multicentre, non-inferiority, randomised controlled trial
Safety and efficacy of inhaled itraconazole in adults with asthma and allergic bronchopulmonary aspergillosis (PUR1900-ABPA): a randomised, double-blind, parallel group, placebo-controlled, multicentre, phase 2 trial
ResearchChest · Nov 21, 2026 · Pulmonology
ResearchChest · Nov 21, 2026 · Pulmonology
Combination acetazolamide and supplemental oxygen markedly improves ventilatory control instability and OSA severity