Vessel-sealing device showed no difference compared to monopolar diathermy in simple mastectomy outcomes
Monopolar diathermy versus a vessel-sealing device for reducing postoperative drain output after simple mastectomy: randomized clinical trial
Clinical question: Does a vessel-sealing device improve perioperative outcomes compared with monopolar diathermy during simple mastectomy?
Why it matters: It is unknown if using a vessel-sealing device instead of monopolar diathermy improves perioperative outcomes like drain output or duration during simple mastectomy.
- Researchers randomized 86 patients to receive either a vessel-sealing device or conventional monopolar diathermy during a simple mastectomy.
- The vessel-sealing device group had a mean drain duration of 8.23 days compared to 7.75 days for monopolar diathermy (P = 0.613).
- Mean total drain output was 572.80 ml for the vessel-sealing device group versus 523.50 ml for the monopolar diathermy group (P = 0.694).
Limitation the authors note: This was a single-centre randomized clinical trial.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Cite this article
Keelan S, Dowling GP, Roche T, Hegarty A, Davey MG, Dhannoon AA. Monopolar diathermy versus a vessel-sealing device for reducing postoperative drain output after simple mastectomy: randomized clinical trial. Br J Surg. 2024. doi:10.1093/bjs/znae029