Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Background
The Aveir leadless pacemaker uses an active fixation method, enabling real-time monitoring of electrical parameters during implantation. However, comprehensive studies on the electrical parameters during this procedure are rare. This study aims to analyze the electrical characteristics to further guide the implantation strategy and improve device stability and safety. Methods
This multicenter retrospective study enrolled 119 patients (mean age, 70.18 years; 59.58% women) who received the Aveir VR leadless pacemaker from November 2024 to May 2025 at 10 centers in China. Intraprocedural variations in commanded electrogram, current of injury (COI), impedance, pacing threshold, and sensing parameters were meticulously documented. Results
Commanded electrogram mapping demonstrated various morphologies (R, RS, QR, QRS, and QS), aiding localization. During fixation, 58.82% of patients exhibited an increased COI from mapping to 0.5 turns, which was associated with reduced short-term pacing thresholds. From 0.5 to 1 turn, 52.94% showed a further increase in COI. Receiver operating characteristic analysis revealed that an impedance increase has predictive value for short-term pacing thresholds, with an area under the receiver operating characteristic curve of 0.634 and a cutoff value of 230 Ω (sensitivity, 0.622; specificity, 0.41). Lead stability showed a moderate correlation with impedance increase (R=0.44, PConclusions
During Aveir implantation, commanded electrogram variations guide site localization. Initial COI increases (0-0.5 turns) are linked to optimal short-term thresholds. Monitoring impedance increase is vital, as a threshold of 230 Ω serves as a key indicator of device stability and fixation quality.
Topics
Spotted a mistake? Suggest a correction
More in Cardiology
Clonal hematopoiesis mutations linked to higher atrial fibrillation risk in cancer populations
Association of Clonal Hematopoiesis Mutations With Incident Atrial Fibrillation: A Systematic Review and Meta-Analysis
Deep learning model detects aortic stenosis from finger-cuff pressure waveforms accurately
Aortic Valve Stenosis Screening Using Deep Learning on Noninvasive Arterial Pressure Waveforms
Spontaneous coronary artery dissection affects younger women more often than expected
Spontaneous coronary artery dissection: implications for women's health, clinical care, and future research
Early lipid-lowering strategies may reduce atherosclerotic cardiovascular disease risk
Lipid-Lowering Strategies in the Era of Early Atherosclerosis Detection: 2026 ACC Scientific Statement: A Report of the American College of Cardiology