ResearchBJA Open · Dec 17, 2025

Noninvasive cardiac index estimation under general anaesthesia: comparison between the VenArt® device and transthoracic echocardiography

Paschoud C, Silvestrini N, Daniels J, Koegel J, Mulin S, Ennahdi-Elidrissi FG, et al
AnesthesiologyAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Background VenArt® is a novel noninvasive cardiac output monitoring device which provides beat-by-beat Fick principle-based measurements of stroke volume and cardiac output. The study aim was to determine the accuracy of this device by comparing it with cardiac output measurements using transthoracic echocardiography in patients undergoing anaesthesia. Methods This prospective observational study included 55 women (ASA physical status classification I-III) undergoing laparoscopic gynaecological procedures. Cardiac output was assessed at five timepoints using the VenArt® device and transthoracic echocardiography. Primary endpoint was the agreement between the two methods regarding cardiac index, evaluated using Bland-Altman analysis to determine bias, precision, mean percentage error, and limits of agreement. Secondary endpoint was the ability of the device to track changes in cardiac index over time compared with echocardiography. Results We analysed 273 pairs of cardiac index values from 55 patients. Bland-Altman analysis showed a bias of 0.02 (95% confidence interval [CI] 0-0.05) L min-1 m-2, with a precision of 0.20 and a mean percentage error of 14.7% (95% CI 13.2-16.2%). Limits of agreement ranged from -0.37 (95% CI -0.41 to -0.33) to 0.41 (95% CI 0.37-0.45) L min-1 m-2. Trending ability demonstrated good agreement: the four-quadrant plot revealed a concordance rate of 95.88%, and the polar plot showed a mean polar angle of 0.75°, with a standard deviation of 13.4° and radial limits of agreement within plus or minus 30°. Conclusions The VenArt® device showed negligible bias and acceptable differences. Trending ability was favourable, with clinically acceptable agreement and high concordance in tracking haemodynamic changes. Clinical trial registration ISRCTN92565809.

DOI: 10.1016/j.bjao.2025.100516 · PubMed: 41503454 · Registered: ISRCTN92565809

Spotted a mistake? Suggest a correction

More in Anesthesiology

Br J Anaesth · Nov 20, 2026 · Research · Anesthesiology

Postoperative delirium accelerates brain ageing by six months

Br J Anaesth · Nov 20, 2026 · Research · Anesthesiology

Vaping-related disorder linked to higher postoperative respiratory complications risk

Br J Anaesth · Nov 20, 2026 · Research · Anesthesiology

Low perfusion index during surgery linked to higher 30-day mortality rates

Br J Anaesth · Nov 17, 2026 · Research · Anesthesiology

Ephedrine and phenylephrine increase human myometrial contractility ex vivo

BJA Open · Nov 13, 2026 · Meta-analysis · Pediatric · Anesthesiology

Effects of propofol- or sevoflurane-containing anaesthesia on auditory brainstem responses in children: a systematic review