Electrically evoked stapedial reflex threshold programming showed similar speech outcomes to behavioral methods in cochlear implant users
Optimizing Cochlear Implant Programming in Single-Sided Deafness and Asymmetric Hearing Loss: A Randomized Comparison of ESRT and Behavioral Methods
Why it matters: Clinicians need to determine if electrically evoked stapedial reflex threshold or behavioral methods are better for programming cochlear implants in patients with single-sided deafness or asymmetric hearing loss.
- Ten participants with single-sided deafness or asymmetric hearing loss were compared using three cochlear implant programming methods in a randomized crossover study.
- Contralateral electrically evoked stapedial reflex threshold programming yielded responses on 100% of active electrodes, while ipsilateral programming yielded responses on 70% of electrodes.
- Speech perception scores were similar across all methods, but 80% of participants preferred the electrically evoked stapedial reflex threshold-based programs.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Cite this article
Holcomb MA, Coto J, Sarangoulis C, Stern T, Prentiss S, Glassman EK. Optimizing Cochlear Implant Programming in Single-Sided Deafness and Asymmetric Hearing Loss: A Randomized Comparison of ESRT and Behavioral Methods. Laryngoscope Investig Otolaryngol. 2026. doi:10.1002/lio2.70601