Phase-resolved MRI shows agreement with standard lung imaging for perfusion and ventilation defects
Phase-resolved MRI for measurement of pulmonary perfusion and ventilation defects in comparison with dynamic contrast-enhanced MRI and 129Xe MRI
Meta-analysis of 10 studiesBenefit found
How does phase-resolved functional lung MRI compare with DCE and 129Xe MRI?
Who
399
399 subjects
Compared
Perfusion defect percentage (QDP): mean difference -2.31% (-8.01% to 3.40%)
Ventilation defect percentage (VDP): mean difference 0.34% (-4.94% to 5.62%)
Main outcome
Perfusion and ventilation defect percentages
Result: QDP correlation 0.65 (95% CI 0.55 to 0.73); VDP correlation 0.72 (95% CI 0.61 to 0.80)
Bottom line: Phase-resolved functional lung MRI shows satisfactory agreement with dynamic contrast-enhanced and 129Xe MRI for assessing lung function.
BMJ Open Respir Res · 2024-08-07 · PROSPERO CRD42023430847
Headline and visual abstract written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Abstract
Introduction
This meta-analysis aims to evaluate the agreement and correlation between phase-resolved functional lung MRI (PREFUL MRI) and dynamic contrast-enhanced (DCE) MRI in evaluating perfusion defect percentage (QDP), as well as the agreement between PREFUL MRI and 129Xe MRI in assessing ventilation defect percentage (VDP). Method
A systematic search was conducted in the Medline, Embase and Cochrane Library databases to identify relevant studies comparing QDP and VDP measured by DCE MRI and 129Xe MRI compared with PREFUL MRI. Meta-analytical techniques were applied to calculate the pooled weighted bias, limits of agreement (LOA) and correlation coefficient. The publication bias was assessed using Egger's regression test, while heterogeneity was assessed using Cochran's Q test and Higgins I2 statistic. Results
A total of 399 subjects from 10 studies were enrolled. The mean difference and LOA were -2.31% (-8.01% to 3.40%) for QDP and 0.34% (-4.94% to 5.62%) for VDP. The pooled correlations (95% CI) were 0.65 (0.55 to 0.73) for QDP and 0.72 (0.61 to 0.80) for VDP. Furthermore, both QDP and VDP showed a negative correlation with forced expiratory volume in 1 s (FEV1). The pooled correlation between QDP and FEV1 was -0.51 (-0.74 to -0.18), as well as between VDP and FEV1 was -0.60 (-0.73 to -0.44). Conclusions
PREFUL MRI is a promising imaging for the assessment of lung function, as it demonstrates satisfactory deviations and LOA when compared with DEC MRI and 129Xe MRI. Prospero registration number
CRD42023430847.
Abstract reproduced under the article’s CC BY NC licence.
Subject terms
Cite this article
Ouyang T, Tang Y, Zhang C, Yang Q. Phase-resolved MRI for measurement of pulmonary perfusion and ventilation defects in comparison with dynamic contrast-enhanced MRI and 129Xe MRI. BMJ Open Respir Res. 2024;11(1). doi:10.1136/bmjresp-2023-002198
Spotted a mistake? Suggest a correction