ResearchAnnals of Intensive Care · Mar 28, 2026
Pulmonary pulsatility quantified by electrical impedance tomography in severe acute respiratory distress syndrome patients undergoing extracorporeal membrane oxygenation support
Leali M, Spinelli E, Giani M, Pavlovsky B, Di Pierro M, Crotti S, et al
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Background
The cardiac-related pulsatility signal from electrical impedance tomography (EIT) correlates with stroke volume in mechanically ventilated patients with acute respiratory distress syndrome (ARDS). However, in swine models, regional pulsatility amplitude was also shown to increase with downstream flow obstruction. We aimed to investigate the relationship between regional pulsatility and pulmonary hemodynamics in a cohort of severe ARDS patients on veno-venous extracorporeal membrane oxygenation (ECMO). Methods
We reanalysed data obtained from 20 ARDS patients receiving ECMO support. EIT was recorded 30 min after adjusting ECMO blood flow to target three ranges of mixed venous oxygen saturation (SvO2) (70-75 %; 75-80 %; >80%), applied in random order. Ventilation was protective with PEEP 15[12-16] cmH2O, Vt 4 ± 1 ml/kg PBW. Quality of EIT tracings allowed for separate modelling of pulsatility and ventilation in 16/20 patients. EIT units were calibrated to millilitres of tidal volume (ml*) using synchronised tidal volume measurements, allowing for between-patient comparisons. Mixed-effects modelling was employed to account for repeated measurements. Results
Across blood flow steps, pulsatility amplitude was directly related to stroke volume (SV) (β = 0.28 (0.06 - 0.5) ml*/mL, p = 0.014) and systolic pulmonary artery pressure (PAPs) (β = 0.47 (0.14 - 0.81) ml*/mmHg, p = 0.008) and inversely related to mixed venous oxygen tension (PvO2) (β = -0.41 (-0.79--0.03) ml*/mmHg, p = 0.039). Changes in pulsatility had an 80% concordance rate with changes in SV and 83% with PAPs. At each ECMO blood flow step, there was a decrease in ventral lung pulsatility during inspiration (p Conclusion
n severe ARDS patients on ECMO, pulsatility amplitude reflects stroke volume changes induced by positive intrathoracic pressures and mixed venous saturation targets. However, downstream flow obstruction appears to be the leading determinant in the dorsal lung and may be useful to monitor right heart loading in patients with ARDS.
DOI: 10.1016/j.aicoj.2026.100058 · PubMed: 41969359
Spotted a mistake? Suggest a correction
More in Critical Care
Meta-analysisCrit Care · Nov 19, 2026 · Critical Care
Discordant evidence on corticosteroids in sepsis: a meta-research study
ResearchPediatricResuscitation · Nov 17, 2026 · Critical Care
Survival after neonatal chest compressions at birth: Temporal trends and prognostic factors in a population-based cohort, Queensland, Australia
ResearchResuscitation · Nov 17, 2026 · Critical Care
Survival after out of hospital cardiac arrest in relation to AED usage in a multi response system including AED equipped drones
ResearchResuscitation · Nov 17, 2026 · Critical Care
Out-of-hospital cardiac arrests due to foreign body airway obstruction: A CanROC registry, retrospective cohort study
ResearchResuscitation · Nov 17, 2026 · Critical Care