Background
The Medtronic Freestyle is widely used for full-root aortic valve replacement, offering excellent hemodynamics. However, structural valve deterioration, including pseudoaneurysm formation, raises concerns. Methods
Our single-center retrospective and prospective cohort study of adult patients who underwent modified Bentall procedures with the Medtronic Freestyle between September 2015 and November 2023 analyzed early and late complications, echocardiographic results, and survival. A cross-sectional computed tomographic angiography (CTA) evaluation assessed postoperative structural abnormalities. Results
Included were 153 patients (median European System for Cardiac Operative Risk Evaluation 2011 revision, 3.4%). The 1- and 5-year actuarial survival rates were 96.7% and 87.2%, respectively. Postoperative mean and maximal transvalvular gradients were 5.8 mm Hg (interquartile range, 4.3-7.6 mm Hg) and 11.1 mm Hg (interquartile range, 8.2-14.9 mm Hg). Of the 98 patients (77.8% of eligible patients) who underwent CTA imaging, 23.5% had structural abnormalities. Pseudoaneurysms were the most common (15.3%) and warranted reintervention in 3 cases. Routine echocardiography failed to detect the pseudoaneurysms. Multivariate analysis identified previous cardiac surgery as an independent predictor of pseudoaneurysm development (odds ratio, 4.32; P = .046). The cumulative risk of CTA-detectable structural abnormalities may reach 22.7% at 5 years. Conclusions
The Medtronic Freestyle displays excellent hemodynamic performance. However, silent structural deterioration is not rare. Routine echocardiography may miss these lesions, supporting the role of systematic postoperative CTA imaging, especially for redo patients. Further studies are needed to guide optimal imaging surveillance protocols.
Topics
Humans; Female; Male; Aortic Valve; Heart Valve Prosthesis Implantation; Retrospective Studies; Heart Valve Prosthesis; Aged; Computed Tomography Angiography; Animals; Bioprosthesis; Postoperative Complications
DOI: 10.1016/j.athoracsur.2026.03.025 · PubMed: 41881409
Spotted a mistake? Suggest a correction
More in Cardiothoracic & Vascular Surgery
ResearchPediatricAnn Thorac Surg · Nov 20, 2026 · Cardiothoracic & Vascular Surgery
One-Year Survival After Hybrid Stage 1 Single Ventricle Palliation - The Impact of Institutional Experience
ResearchJ Vasc Surg · Nov 16, 2026 · Cardiothoracic & Vascular Surgery
Incidence of Aortic Vascular Graft and Endograft Infection in Patients with Bloodstream Infection - A Population-Based Study
ResearchJ Vasc Surg · Nov 14, 2026 · Cardiothoracic & Vascular Surgery
Meta-analysisEur J Vasc Endovasc Surg · Nov 13, 2026 · Cardiothoracic & Vascular Surgery
Single vs. Dual Antiplatelet Therapy after Fenestrated and Branched Endovascular Aortic Repair: A Systematic Review and Meta-analysis
ResearchJ Thorac Cardiovasc Surg · Nov 12, 2026 · Cardiothoracic & Vascular Surgery
Early versus delayed surgery to prevent hemorrhagic transformation in patients with septic cerebral emboli from infective endocarditis