Test-Retest Repeatability of a Novel Photopic Microperimetry Approach in Patients with Geographic Atrophy
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Purpose
To evaluate test-retest repeatability of novel focal photopic microperimetry approach targeting central 1 mm of the fovea and explore its relationships with geographic atrophy (GA) lesion area and other visual function metrics. Design
A prospective, single arm, observational study. Subjects
Participants with nonsubfoveal GA as determined by spectral-domain OCT and fundus autofluorescence. Methods
Twelve participants (17 eyes) performed focal photopic tests using a 45-loci focal grid (0.5° spacing; Goldmann II stimuli) on Nidek MP-3, along with other visual function assessments. To assess test-retest repeatability of MP-3, 2 tests up to 4 weeks apart were performed. The interval was short relative to expected GA progression, and thus measurements are interpreted as reflecting test-retest repeatability. Main outcome measures
Coefficient of repeatability (CoR), intraclass correlation coefficient (ICC), and Bland-Altman were analyzed for mean sensitivity (MS) and point-wise sensitivity (PWS). Correlations of adjusted MS (excluding floor effect) with GA lesion size, best-corrected visual acuity (BCVA), low-luminance visual acuity (LLVA), and contrast sensitivity were also analyzed. Results
Focal photopic MP-3 showed good test-retest repeatability with CoR of MS at 2.42 dB (intraclass correlation coefficient [ICC] = 0.98) and CoR of PWS at 8.2 dB (ICC = 0.89). Excluding the floor effect improved the CoR of MS and PWS to 2.03 dB and 7.06 dB, respectively. Mean test duration was 4.33 minutes per eye with fixation stability >89%. Structure-function analysis showed weak correlation between GA lesion size and adjusted MS (P = 0.05; R 2 = 0.44). Adjusted MS showed a significant correlation with LLVA (P = 0.001) and low-luminance quick contrast sensitivity function (qCSF) (P = 0.01), a weak association with qCSF (P = 0.05), but no correlation with BCVA. Conclusions
Focal photopic microperimetry using Goldmann II stimuli provides a highly repeatable measure of central cone-mediated function in patients with nonsubfoveal GA. Mean sensitivity demonstrated high repeatability and supports its use as a robust, clinically relevant endpoint for longitudinal and interventional studies, while PWS may provide spatial information for deeper structure-function analyses. Weak associations with GA lesion size and BCVA suggest that conventional measures may underestimate visual impairment, highlighting the value of focal photopic microperimetry within central 1 mm region as a sensitive, patient-centered outcome in age-related macular degeneration. Financial disclosures
Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.
Headline, design, result, key points, caveat and citation as plain text.
Cite this article
Tandon NR, Vasquez E, Csaky KG. Test-Retest Repeatability of a Novel Photopic Microperimetry Approach in Patients with Geographic Atrophy. Ophthalmol Sci. 2026;6(8):101267. doi:10.1016/j.xops.2026.101267
Spotted a mistake? Suggest a correction