Use of Fibre Optic RealShape Technology during Complex Endovascular Aortic Repair: Results of a Transatlantic Registry
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Objective
This study describes the first transatlantic multicentre results of catheterisation success rates and radiation reducing effect of Fibre Optic RealShape (FORS) technology in complex endovascular aortic repair (EVAR) procedures. Methods
Data were derived from the transatlantic, multicenter, prospective observational FORS Learn Registry, executed at seven high-volume international centers specialized in complex endovascular aortic interventions. All target vessel catheterisations during complex EVAR procedures were analysed per protocol, focusing on the technical success of FORS and the impact of FORS on navigation time and radiation exposure, measured by fluoroscopy time (FT), fluoroscopy rate (FR), cumulative air kerma (CAK), and kerma area product (KAP) by comparing successful FORS with conventional catheterisations. Results
During 313 complex EVARs, 812 FORS attempts at 1 037 visceral target catheterisations resulted in an overall technical success of 62.8% (fenestrations 66.2% vs. branches 54.1%; p = .001). Successful FORS catheterisation was significantly faster through fenestrations (5.0 minutes [interquartile range {IQR} 3.0, 7.0] vs. 6.5 minutes [IQR 3.0, 13.0]; p Conclusion
This multicentre study showed the potential of FORS technology to reduce radiation exposure during visceral target catheterisation in complex EVARs with reasonable technical success.
Cite this article
Lemmens CC, Klaassen J, Beck AW, Kölbel T, Schanzer A, Schermerhorn ML. Use of Fibre Optic RealShape Technology during Complex Endovascular Aortic Repair: Results of a Transatlantic Registry. Eur J Vasc Endovasc Surg. 2026. doi:10.1016/j.ejvs.2026.04.029