Validation of Automated Fixed Force Applanation Tonometry

Wen JC, Vora R, Jiang H, McKinnon S, Muir KW, Tantum SL
OphthalmologyAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Purpose To validate fixed force applanation tonometry against traditional Goldmann applanation tonometry (GAT) and compare interuser reproducibility. Design Prospective cross-sectional pilot study. Subjects Patients presenting for their routine eye care at an academic glaucoma practice. Methods A prototype adaptor containing a camera was attached to the Goldmann tonometer of a standard slitlamp microscope set to a fixed force in conjunction with a custom tonometer tip. Recorded applanation mires were converted to an intraocular pressure (IOP) using the Imbert-Fick formula. Standard GAT was also performed. Clinical GAT, research GAT, and prototype IOP measurements among different users were also completed. Main outcome measures Goldmann applanation tonometry and prototype IOPs were compared in a pilot validation study, and interuser reproducibility was assessed using the 2 devices. Results In the validation study, 100 eyes from 50 subjects were included. The mean difference between the prototype and standard GAT measurements was +1.0 mmHg, with 95% limits of agreement (LoA) of -3.4 to 5.5 mmHg. The mean absolute difference (± standard deviation) between the prototype IOP and the standard GAT measurement was 1.9 ± 1.5 mmHg. In the reproducibility study, 42 eyes from 21 subjects were included. The mean difference between the research technician prototype IOP and the ophthalmologist prototype IOP measurements was 0.01 mmHg with LoA of -2.5 to 2.4 mmHg. The mean absolute difference between users was significantly lower when using the prototype compared with standard GAT (0.9 ± 1.2 mmHg vs. 1.6 ± 1.1 mmHg, P = 0.003). Conclusions Preliminary measurements using a fixed force applanation tonometry method demonstrated comparable results with standard GAT with improved interuser reproducibility. Financial disclosures The authors have no proprietary or commercial interest in any materials discussed in this article.

DOI: 10.1016/j.xops.2026.101370 · PubMed: 42763533

Spotted a mistake? Suggest a correction

More in Ophthalmology

ResearchPediatricAm J Ophthalmol · Nov 20, 2026 · Ophthalmology

Allergic conjunctivitis severity linked to meibomian gland loss in children

Association of allergic conjunctivitis severity with meibomian gland abnormalities in children: a cross-sectional study

Randomized trialVisual abstractPediatricJAMA Ophthalmol · Nov 19, 2026 · Ophthalmology

Higher concentration atropine drops slowed myopia progression in children

Low-Concentration Atropine Eye Drops in Children With Myopia and Intermittent Exotropia: The AMIXT2 Randomized Clinical Trial

ResearchAm J Ophthalmol · Nov 19, 2026 · Ophthalmology

Wildfire smoke exposure linked to Leber's Hereditary Optic Neuropathy conversion

Smoke, Not Smoking, as a Possible Risk Factor Triggering Conversion in Leber's Hereditary Optic Neuropathy: A Retrospective Case Series

ResearchOphthalmology · Nov 18, 2026 · Ophthalmology

Enucleations decreased and globe-sparing radiation increased for uveal melanoma

Trends and Variation in Uveal Melanoma Management in the National Cancer Database, 2004-2023

ResearchAm J Ophthalmol · Nov 18, 2026 · Ophthalmology

Static perimetry thresholds correlate with kinetic perimetry in retinitis pigmentosa patients

Developing Static Perimetry Thresholds to Assess Visual Field Loss Severity in Patients with Retinitis Pigmentosa