Validation of Non-Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry

Ensing AE, Mueller L, Lee DS, Kallogjeri D, Lieu JEC
ENT (Otolaryngology)PediatricsPediatric
Read the full paper free at PubMed Central ↗

Abstract

Objectives Portable, tablet-based audiometry has been proposed as a diagnostic tool for patients unable to access traditional sound booth audiometry. However, the validity of tablet-based audiometry for diagnostic use in pediatric patients has not been established. The objective of this study is to validate self-administered tablet-based diagnostic audiometry in a pediatric clinic setting. Methods Children 4 to 17.9 years-old were recruited from a pediatric audiology clinic. Tablet audiometry was performed with tablet-based software, SHOEBOX Audiometry (Ottawa, ON) at six frequencies (250-8000 Hz). Hearing thresholds from traditional pure tone audiometry in a sound booth were used as a gold-standard comparison. Hearing loss was defined as ≥ 1 hearing threshold of > 20 dB. Results 120 children underwent tablet-audiometry, though not all were able to complete the test. In 948 individual frequencies tested, the mean threshold difference was -0.95 dB (95% CI -1.5 to -0.42 dB) between traditional and tablet audiogram. Most tablet-assessed thresholds (92%, n = 873) were within 10 dB of corresponding traditional audiometry thresholds. Tablet audiometry demonstrated 92% sensitivity (95% CI 84%-97%) and 85% specificity (95% CI 77%-92%) for diagnosis of hearing loss. Test-retest reliability of tablet audiometry was high, with 93% (55/59) of test-retest hearing threshold differences falling within ±5 dB. Conclusion Tablet audiometry is valid and reliable for children aged 5-17 years who can successfully complete testing in a non-soundproof clinic setting. It may serve as a tool for remote hearing testing, with potential applications in schools, clinics, and research.

Topics

Humans; Child; Child, Preschool; Adolescent; Female; Male; Audiometry, Pure-Tone; Computers, Handheld; Hearing Loss; Reproducibility of Results; Auditory Acuity; Auditory Threshold

DOI: 10.1002/lary.70514 · PubMed: 41856760

Cite this article

Ensing AE, Mueller L, Lee DS, Kallogjeri D, Lieu JEC. Validation of Non-Soundproof Tablet Audiometry in Pediatrics: Comparison With Traditional Audiometry. Laryngoscope. 2026. doi:10.1002/lary.70514

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Otolaryngol Head Neck Surg · May 20, 2026 · Research · Pediatric · ENT (Otolaryngology), Pediatrics

Longitudinal Evaluation of Otoacoustic Emissions as a Screening Tool for High-Frequency Hearing Loss in Adolescents

Laryngoscope · Jan 21, 2026 · Research · ENT (Otolaryngology)

Robotic-Assisted Electrode Array Insertion Improves Stability of Acoustic Hearing Thresholds

J Gerontol A · Sep 4, 2026 · Research · Geriatrics, ENT (Otolaryngology)

Association between midlife macronutrient intake and late-life hearing loss: the Atherosclerosis Risk in Communities Study

BMC Prim Care · May 20, 2026 · Research · Family Medicine & Primary Care, ENT (Otolaryngology)

Determinants of poor speech discrimination in a primary care otolaryngology population: a retrospective observational study

J Otolaryngol Head Neck Surg · Nov 20, 2026 · Research · ENT (Otolaryngology)

Hearing improvement at day 6 was associated with better final hearing recovery in patients with sudden hearing loss