Vital sign parameters are not associated with electrical capture during transcutaneous pacing
Kimbrell JM, Kreinbrook JA, Smida T, Gan GH, Odom W, Shukis M, et al
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Background
Failure to achieve electrical capture during transcutaneous pacing (TCP) is common and underrecognized. We aimed to identify whether pulse oximetry plethysmograph waveform synchrony, end-tidal capnometry increases, and systolic blood pressure increases were associated with sustained electrical capture during prehospital TCP. Methods
This secondary analysis of a multi-center retrospective observational study enrolled patients from four emergency medical service agencies. Sustained electrical capture was defined as a wide QRS and T wave for ≥90% of ongoing TCP time. Pulse oximetry plethysmograph wave synchronization with the TCP rate and vital sign values were obtained from cardiac monitor/defibrillator files. A diagnostic framework was used to evaluate plethysmograph synchronization. We used linear regression modeling to detect whether increases in end-tidal carbon dioxide (ETCO2) and blood pressure measurements were associated with sustained electrical capture. Results
Of 253 patients, 29 (11.5%) had sustained electrical capture. An increase in SBP had a sensitivity of 75.0% (42.8, 94.5) and a specificity of 40.0% (29.8, 50.9) for the outcome of sustained electrical capture. An increase in ETCO2 had a sensitivity of 41.7% (15.2, 72.3) and a specificity of 64.4% (53.7, 74.3). Linear regression modeling did not suggest that sustained electrical capture was associated with post-TCP SBP (-2.5 mmHg (-27.3, +22.4)) or post-TCP ETCO2 (2.6 mmHg (-2.9, +8.1)). Plethysmograph synchronization had a sensitivity of 58.8% (32.9, 81.6) and a specificity of 81.1% (73.2, 87.5). Conclusions
These findings suggest that plethysmograph waveforms, capnometry, and blood pressure monitoring alone are not reliable indicators of electrical capture during TCP performed by paramedics.
DOI: 10.1016/j.ajem.2026.07.050 · PubMed: 42600263
Spotted a mistake? Suggest a correction
More in Emergency Medicine
Meta-analysisPLOS Med · Nov 19, 2026 · Infectious Disease, Nephrology, Emergency Medicine
Acute kidney injury and death in severe Plasmodium falciparum malaria: An individual patient data meta-analysis
ResearchBMJ Med · Nov 18, 2026 · Emergency Medicine
Association between direct admission of patients with severe injuries to a level 1 trauma centre and mortality: nationwide cohort study
ResearchAm J Emerg Med · Nov 13, 2026 · Emergency Medicine
ED red blood cell transfusion in severe anemia: stay and recurrent use-a retrospective cohort study
ResearchAm J Emerg Med · Nov 12, 2026 · Emergency Medicine
Serum lactate distribution across neurologic emergencies: implications for seizure identification
ResearchAm J Emerg Med · Nov 12, 2026 · Emergency Medicine
Longer wait times after ED-initiated buprenorphine associated with lower appointment attendance