Randomized trialClinical Chemistry · Dec 11, 2022

Within-Person and Between-Sensor Variability in Continuous Glucose Monitoring Metrics

Selvin E, Wang D, Rooney MR, Fang M, Echouffo-Tcheugui JB, Zeger S, et al
Pathology & Laboratory MedicineEndocrinology & DiabetesAdult
Read the full paper free at PubMed Central ↗

Abstract

Background The within-person and between-sensor variability of metrics from different interstitial continuous glucose monitoring (CGM) sensors in adults with type 2 diabetes not taking insulin is unclear. Methods Secondary analysis of data from 172 participants from the Hyperglycemic Profiles in Obstructive Sleep Apnea randomized clinical trial. Participants simultaneously wore Dexcom G4 and Abbott Libre Pro CGM sensors for up to 2 weeks at baseline and again at the 3-month follow-up visit. Results At baseline (up to 2 weeks of CGM), mean glucose for both the Abbott and Dexcom sensors was approximately 150 mg/dL (8.3 mmol/L) and time in range (70180 mg/dL [3.910.0 mmol/L]) was just below 80. When comparing the same sensor at 2 different time points (two 2-week periods, 3 months apart), the within-person coefficient of variation (CVw) in mean glucose was 17.4 (Abbott) and 14.2 (Dexcom). CVw for percent time in range: 20.1 (Abbott) and 18.6 (Dexcom). At baseline, the Pearson correlation of mean glucose from the 2 sensors worn simultaneously was r 0.86, root mean squared error (RMSE), 13 mg/dL (0.7 mmol/L); for time in range, r 0.88, RMSE, 8 percentage points. Conclusions Substantial variation was observed within sensors over time and across 2 different sensors worn simultaneously on the same individuals. Clinicians should be aware of this variability when using CGM technology to make clinical decisions.ClinicalTrials.gov Identifier: NCT02454153.

Topics

Adult; Humans; Blood Glucose; Diabetes Mellitus, Type 1; Blood Glucose Self-Monitoring; Diabetes Mellitus, Type 2; Insulin

DOI: 10.1093/clinchem/hvac192 · PubMed: 36495162 · Registered: NCT02454153

Cite this article

Selvin E, Wang D, Rooney MR, Fang M, Echouffo-Tcheugui JB, Zeger S. Within-Person and Between-Sensor Variability in Continuous Glucose Monitoring Metrics. Clin Chem. 2022. doi:10.1093/clinchem/hvac192

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top studies every Monday

The week’s key trials and guidelines in your specialties, plus any topics you follow. By email, free.

Get the weekly digest

Similar articles

Review · Diabetes Obes Metab · Jan 27, 2026 · Endocrinology & Diabetes

Rethinking control-IQ+ technology: Simple strategies for easy optimization

Guideline · Diabetologia · Mar 12, 2026 · Endocrinology & Diabetes

Diabetes experts call for personalized technology use to address access gaps and digital divides

Randomized trial · Diabetes Obes Metab · May 25, 2022 · Endocrinology & Diabetes, Gastroenterology & Hepatology

Reduced hypoglycaemia using liver-targeted insulin in individuals with type 1 diabetes

Guideline · Lancet Diabetes Endocrinol · Dec 17, 2025 · Endocrinology & Diabetes, Obstetrics & Gynecology

International experts recommend glucose monitoring and automated insulin delivery for pregnant women with diabetes

Randomized trial · Diabetes Care · Mar 16, 2023 · Endocrinology & Diabetes

Novel Once-Weekly Basal Insulin Fc Achieved Similar Glycemic Control With a Safety Profile Comparable to Insulin Degludec in Patients With Type 1 Diabetes