Visual abstractsVisuals
ResearchInsights into Imaging · Aug 26, 2026

AI performance varies considerably across mammography devices: a multi-site and multi-vendor retrospective study

Blum M, Morant R, Eichenberger A, Geissler A, Subelack J, Vogel J, et al
Radiology & ImagingAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Objectives To evaluate an artificial intelligence (AI) algorithm on three different mammography devices and assess screening performance using general and device-specific thresholds. Materials and methods This retrospective study analyzed 22,673 AI-assessed mammographies of women aged 50-69 collected in 2022-2023 from three mammography devices across six screening sites within a Swiss Mammography Screening Program. 115 screen-detected and 15 1-year interval breast cancers were included in the analyses. A commercially available AI algorithm assessed each mammography with a numeric case score. Optimal thresholds were determined across all devices and for each device separately using the Youden-Index. Screening performance was assessed using sensitivity, specificity, balanced accuracy, and cancer detection rate (CDR). Results Standard double-reading had a CDR of 5.07 (95% CI: 4.19-6.09). Using AI with a general threshold achieved a CDR of 4.81 (95% CI: 3.95-5.80), which was enhanced using device-specific thresholds to 4.90 (95% CI: 4.03-5.89). Workload for radiologists in consensus conferences was lowered by nearly one-third when using device-specific thresholds compared to a general threshold (p Conclusion AI-based performance varied considerably among the three mammography devices, emphasizing the need for device-specific thresholds or even the withdrawal of AI use on certain devices. Awareness of different case distribution among radiologists should be raised to mitigate the risk of interpretation biases and reduced diagnostic accuracy. Lastly, training datasets must be representative not only of the target population but also of the different screening devices used in practice. Key points Question AI performance may vary across mammography devices, raising uncertainty about device-independent thresholds. Findings AI scores differed considerably across devices, indicating the need for device-specific thresholds, but performance on one device remained poor. Critical relevance The performance of an AI algorithm in mammography screening depends on adequately defined decision thresholds, and radiologists must be aware that AI scores can vary across mammography devices to avoid interpretation biases and to maintain diagnostic accuracy.

DOI: 10.1186/s13244-026-02383-5 · PubMed: 42645664

Spotted a mistake? Suggest a correction

More in Radiology & Imaging

ResearchInsights Imaging · Nov 20, 2026 · Radiology & Imaging

Integrated prognostic system predicts disease-free survival in pancreatic cancer patients

Enhanced prognosis stratification following surgery for pancreatic ductal adenocarcinoma to optimize clinical decisions by reflecting tumor heterogeneity and microenvironment

ResearchInsights Imaging · Nov 19, 2026 · Radiology & Imaging

Dual-phase photon-counting CT localized parathyroid adenomas with 67.3% to 83.7% sensitivity

Diagnostic performance of dual-phase photon-counting detector for localization of parathyroid adenomas after inconclusive first-line imaging

ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging

Advancing placenta accreta spectrum diagnosis: is 3-T MRI or combined MRI with ultrasound the answer?

ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging

Classic imaging signs aid diagnosis in head and neck radiology

Classic signs in head and neck imaging-a quick guide for the radiology resident

ResearchInsights Imaging · Nov 15, 2026 · Radiology & Imaging

Automated carotid plaque quantification helps identify obstructive coronary artery disease

Carotid CTA with automated plaque quantification for risk stratification of coexisting obstructive coronary artery disease