Two automated software tools showed limited agreement when assessing breast positioning quality in screening
Comparison of two automated PGMI-based software tools for breast positioning quality assessment in DBT screening
Why it matters: The agreement between different automated tools for assessing breast positioning quality in digital breast tomosynthesis remains uncertain.
- Researchers compared two automated software tools, SW1 and SW2, using 3,112 views from 778 women in a retrospective study.
- The overall agreement between the two systems was limited at 48.2% with a quadratic weighted kappa of 0.044.
- SW2 was associated with more Inadequate ratings (10.9%) compared to SW1 (0.7%), while SW1 had more Perfect ratings (21.9%) than SW2 (12.9%).
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Topics
Cite this article
Gennaro G, Trazzi S, Bordon V, Galli V. Comparison of two automated PGMI-based software tools for breast positioning quality assessment in DBT screening. Eur Radiol Exp. 2026. doi:10.1186/s41747-026-00825-9
Spotted a mistake? Suggest a correction