ResearchJournal of Nuclear Medicine · Feb 2, 2026

Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study

Chen Y, Mintzer RA, Cho S, Pocci D, Hamdi M, Komarov S, et al
Radiology & ImagingAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Augmented whole-body scanning via magnifying PET (AWSM-PET) is a technology that aims to enhance the image resolution and sensitivity of the clinical PET/CT scanner for whole-body imaging. This study presents the system design, performance evaluation, and results from phantoms and initial human imaging studies. Methods: The AWSM-PET system integrates 2 high-resolution detector panels ("outserts") with a Biograph Vision PET/CT scanner (Siemens Healthineers). Positioned outside the scanner's axial field of view, each panel consists of 32 detector modules, each containing a 30 × 30 lutetium oxyorthosilicate crystal array (0.97 × 0.97 × 10 mm³ per crystal) at 1.05-mm pitch. These detectors are packaged and read out using the scanner's detector electronics assembly. Customized firmware and software were developed to establish the additional coincidence detection among the outserts and the scanner. The system acquires data simultaneously during a whole-body scan using continuous bed motion. List-mode-based image reconstruction software has been developed for AWSM-PET system using the continuous-bed-motion protocol. Performance was evaluated through sensitivity measurements, imaging of a mini-resolution phantom containing multiple groups of hot spheres, a National Electrical Manufacturers Association Image Quality (NEMA IQ) phantom study with custom small-diameter tumor inserts, and an initial human study. Results: The outsert detectors achieved an energy resolution of 11% at 511 keV. The coincidence resolving time between the outserts and between outserts and scanner was 183 and 211 ps full width at half maximum, respectively. Sensitivity increased by up to 18.4%, depending on the source location. Mini-resolution phantom images demonstrated higher peak-to-valley ratios for spheres with a diameter of 6 mm or less compared with native scanner images. In the NEMA IQ study, AWSM-PET achieved a higher contrast recovery ratio (CRC) for the smallest spheric lesion (diameter, 4.88 mm) across all CRC metrics (maximum, peak, and mean) but no improvement for larger lesions compared with high-resolution Biograph Vision images. Initial human imaging confirmed the system's compatibility with clinical workflows, achieving improved resolution in high-count regions, such as the brain. However, increased noise in low-count regions (e.g., abdomen) was observed and requires further improvement. Conclusion: The prototype AWSM-PET system demonstrated enhanced image resolution and sensitivity of a clinical PET/CT scanner without negatively impacting its performance, showing improved CRC and spatial resolution for small lesions in phantom and human imaging.

Topics

Humans; Whole Body Imaging; Phantoms, Imaging; Positron-Emission Tomography; Equipment Design; Positron Emission Tomography Computed Tomography; Image Processing, Computer-Assisted; Pilot Projects

DOI: 10.2967/jnumed.125.270117 · PubMed: 41344860

Spotted a mistake? Suggest a correction

More in Radiology & Imaging

Insights Imaging · Nov 20, 2026 · Research · Radiology & Imaging

Integrated system predicts disease-free survival in pancreatic cancer patients better

Insights Imaging · Nov 19, 2026 · Research · Radiology & Imaging

Dual-phase photon-counting detector localized parathyroid adenomas with 67.3% to 83.7% sensitivity

Insights Imaging · Nov 18, 2026 · Research · Radiology & Imaging

MRI detected placenta accreta spectrum with 100 percent sensitivity

Insights Imaging · Nov 18, 2026 · Research · Radiology & Imaging

Classic imaging signs aid diagnosis in head and neck radiology

Insights Imaging · Nov 15, 2026 · Research · Radiology & Imaging

Carotid plaque quantification helps identify patients with obstructive coronary artery disease