Diagnostic Accuracy of Assays Using Point-of-Care Testing or Dried Blood Spot Samples for the Determination of Hepatitis C Virus RNA: A Systematic Review
Systematic review and meta-analysisBenefit found
What is the diagnostic accuracy of point-of-care and dried blood spot hepatitis C virus RNA testing?
Who
Hepatitis C virus RNA samples
Compared
Xpert HCV Viral Load Fingerstick: 99% (95% CI, 97%-99%) sensitivity
Dried blood spot: 97% (95% CI, 95%-98%) sensitivity
Main outcome
Diagnostic accuracy for hepatitis C virus RNA detection and quantification
Bottom line: Point-of-care finger-stick and dried blood spot assays show excellent diagnostic accuracy for hepatitis C virus RNA.
J Infect Dis · 2022-02-13Download image
Clinical question: What is the diagnostic accuracy of point-of-care and dried blood spot hepatitis C virus RNA testing?
Abstract
Background
Finger-stick point-of-care and dried blood spot (DBS) hepatitis C virus (HCV) RNA testing increases testing uptake and linkage to care. This systematic review evaluated the diagnostic accuracy of point-of-care testing and DBS to detect HCV RNA. Methods
Bibliographic databases and conference presentations were searched for eligible studies. Meta-analysis was used to pool estimates. Results
Of 359 articles identified, 43 studies were eligible and included. When comparing the Xpert HCV Viral Load Fingerstick assay to venous blood samples (7 studies with 987 samples), the sensitivity and specificity for HCV RNA detection was 99% (95% confidence interval [CI], 97%-99%) and 99% (95% CI, 94%-100%) and for HCV RNA quantification was 100% (95% CI, 93%-100%) and 100% (95% CI, 94%-100%). The proportion of invalid results following Xpert HCV Viral Load Fingerstick testing was 6% (95% CI, 3%-11%). When comparing DBS to venous blood samples (28 studies with 3988 samples) the sensitivity and specificity for HCV RNA detection was 97% (95% CI, 95%-98%) and 100% (95% CI, 98%-100%) and for HCV RNA quantification was 98% (95% CI, 96%-99%) and 100% (95% CI, 95%-100%). Conclusions
Excellent diagnostic accuracy was observed across assays for detection of HCV RNA from finger-stick and DBS samples. The proportion of invalid results following Xpert HCV Viral Load Fingerstick testing highlights the importance of operator training and quality assurance programs.