Diagnostic Accuracy of Assays Using Point-of-Care Testing or Dried Blood Spot Samples for the Determination of Hepatitis C Virus RNA: A Systematic Review
Key result: Finger-stick and dried blood spot tests accurately detect hepatitis C virus RNA
Systematic review and meta-analysisKey finding
What is the diagnostic accuracy of point-of-care and dried blood spot hepatitis C virus RNA testing?
Who
Hepatitis C virus RNA samples
Compared
Xpert HCV Viral Load Fingerstick: 99% (95% CI, 97%-99%) sensitivity
Dried blood spot: 97% (95% CI, 95%-98%) sensitivity
Main outcome
Diagnostic accuracy for hepatitis C virus RNA detection and quantification
Bottom line: Point-of-care finger-stick and dried blood spot assays show excellent diagnostic accuracy for hepatitis C virus RNA.
J Infect Dis · 2022-02-13 · Save PNG
Key points
- This systematic review analyzed 43 studies comparing finger-stick and dried blood spot samples to venous blood for hepatitis C virus RNA.
- The Xpert finger-stick assay showed 99% sensitivity and 99% specificity for detection, while dried blood spot samples showed 97% sensitivity and 100% specificity.
- The Xpert finger-stick assay produced 6% invalid results, highlighting the need for operator training and quality assurance programs.
Discussion prompts
- Would these patients look like yours? Who was left out?
- Is the main outcome one that matters to patients?
- Is the difference large enough to change what you do?
Cite: Catlett B, Hajarizadeh B, Cunningham E, et al. Diagnostic Accuracy of Assays Using Point-of-Care Testing or Dried Blood Spot Samples for the Determination of Hepatitis C Virus RNA: A Systematic Review. J Infect Dis. 2022;226(6):1005-1021. doi:10.1093/infdis/jiac049
Free full text: J Infect Dis website · Summary and visual abstract: nowallmed.com/a/diagnostic-accuracy-of-assays-using-point-of-care-testing-or-dried-1uezs4e/ (AI-assisted from the abstract; check the paper before relying on it).