Discrepancies Between Point of Care and Central Laboratory Sodium and Potassium Measurements in ICU: Analytical Biases and Physician Awareness

Nakhil N, Najem S, Driss R, Chaabouni T, Dufour N
Critical CareAdult
Read the full paper free at Europe PMC ↗
Top Critical Care studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Abstract

Objectives Blood electrolyte panels, especially sodium ion (Na + ) and potassium ion (K + ), are fundamental and routine tests. In critical care settings, these measurements are typically performed by central laboratories using indirect potentiometry (IP) or via point-of-care blood gas analyzers using direct potentiometry (DP). Discrepancies between the two methods exist, and clinicians often have limited knowledge regarding the strengths and weaknesses of each technique. We wanted to quantify these observations. Design We compared prospectively Na + and K + measurements obtained from the same blood draw using both IP and DP (including analyses of correlation, agreement, and discrepancies). We also conducted a brief survey among physicians to assess their knowledge regarding IP and DP. Settings A tertiary hospital. Patients A total of 501 paired measurements were prospectively collected from an ICU population. Interventions None. Measurements and main results We assessed the degree of agreement between the two methods. We also examined the impact of proteinemia and hemolysis on Na + and K + values, respectively. Only 31.1% of the 103 responding physicians were aware of the analytical bias in Na + measurements obtained by IP, and 45.6% considered K + measurements from DP to be as reliable or more reliable than those obtained by IP. The agreement between the two methods was moderate (Lin's concordance correlation coefficient: 0.90 for Na + and 0.93 for K + ). The 95% limit of agreement for Na + was particularly large (10.48 mmol/L). The divergence between the two methods, defined as a discrepancy in classification (within, below, or above the normal range), occurred in approximately 10% of cases, for both Na + and K + . Conclusions IP and DP were not interchangeable, each exhibiting distinct strengths and limitations. Enhancing physician awareness of the differences between these methods could improve the quality of care.

Abstract reproduced under the article’s CC BY-NC-ND licence.

Subject terms

Humans; Potassium; Sodium; Potentiometry; Prospective Studies; Intensive Care Units; Point-of-Care Systems

DOI: 10.1097/CCM.0000000000007039 · PubMed: 41823506

Cite this article

Nakhil N, Najem S, Driss R, Chaabouni T, Dufour N. Discrepancies Between Point of Care and Central Laboratory Sodium and Potassium Measurements in ICU: Analytical Biases and Physician Awareness. Crit Care Med. 2026. doi:10.1097/CCM.0000000000007039

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Review · Crit Care · Apr 17, 2026 · Critical Care

EEG for bedside monitoring: the intensivist's point of view

Research · Crit Care · Mar 1, 2026 · Critical Care

Derivation of a pragmatic three-view protocol for point-of-care transesophageal echocardiography in the cardiovascular ICU: a real-world cohort study

Meta-analysis · Crit Care · Jul 19, 2026 · Critical Care

Sodium bicarbonate reduced kidney dialysis use but did not lower overall 90-day mortality

Meta-analysis · Resuscitation · Jan 4, 2025 · Critical Care

Insulin with glucose and salbutamol were associated with lower potassium levels in hyperkalaemia

Guideline · Crit Care · Dec 24, 2020 · Critical Care, Infectious Disease

International experts suggest point-of-care ultrasound for managing respiratory and cardiac aspects of COVID-19