Does transthoracic impedance influence termination and recurrence of ventricular fibrillation during impedance-compensated shocks for out-of-hospital cardiac arrest? A retrospective study
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Aim
To examine the association between transthoracic impedance (TTI) and termination of ventricular fibrillation or pulseless ventricular tachycardia (VF/pVT) and recurrent VF/pVT in out-of-hospital cardiac arrest (OHCA). Methods
We conducted a retrospective observational study of non-traumatic OHCA in patients aged ≥16 years with an initial shockable rhythm between January 2019 and October 2022. Termination of VF/pVT was defined as the absence of a shockable rhythm within 5 s following shock delivery, resulting in asystole or an organised rhythm. The relationship between TTI and VF/pVT termination and recurrence were assessed using mixed-effects logistic regression models applied to shock level data with a random intercept for clustering by patient. Results
Between 2019 and 2022, 1880 cases met the inclusion criteria, comprising 8333 analysable shocks. The unadjusted probability of VF/pVT termination declined at TTI values ≥100 Ω, whereas no clear relationship was observed at TTI <100 Ω. In adjusted analyses, TTI <100 Ω was not associated with VF/pVT termination (adjusted odds ratio [OR] per 1-Ω increase 0.999; 95% CI 0.993-1.006; p = 0.860), while TTI ≥100 Ω was associated with lower odds of termination (adjusted OR per 1-Ω increase 0.994; 95% CI 0.988-0.999; p = 0.016). Among shocks that resulted in VF/pVT termination, TTI was not associated with recurrent VF/pVT (adjusted OR per 1-Ω increase 1.003; 95% CI 0.999-1.006; p = 0.082). Conclusion
Despite impedance-compensated defibrillation, TTI ≥100 Ω remains associated with reduced likelihood of VF/pVT termination in OHCA. In contrast, TTI does not appear to influence the risk of recurrent VF/pVT following successful termination.
Subject terms
Headline, design, result, key points, caveat and citation as plain text.
Cite this article
Alhenaki A, Alqudah Z, Kennett T, Dantanarayana A, Williams B, Nehme Z. Does transthoracic impedance influence termination and recurrence of ventricular fibrillation during impedance-compensated shocks for out-of-hospital cardiac arrest? A retrospective study. Resuscitation. 2026;225:111151. doi:10.1016/j.resuscitation.2026.111151
Spotted a mistake? Suggest a correction