Dual-phase systolic and diastolic cardiac diffusion tensor imaging using higher order motion compensation spin echo sequences at 300 mT/m gradient strength

Afzali M, Teh I, Coveney S, Mueller L, Watson I, Lwin M, et al
Radiology & ImagingAdult
Read the full paper free at the publisher ↗

Abstract

Background Cardiac diffusion tensor imaging (cDTI) provides unique insights into myocardial microstructure and its dynamic changes during the cardiac cycle. However, motion sensitivity remains a major obstacle for spin echo (SE)-based approaches, restricting most implementations to a relatively narrow low-motion window near peak systole. Purpose To establish whether higher-order motion-compensated (up to 6th order) diffusion gradients beyond standard second order motion compensation (M2) on a magnetic resonance imaging (MRI) system with a maximum gradient strength of 300 mT/m improve SE-based cDTI techniques for both systolic and diastolic acquisitions. Methods Ten healthy volunteers underwent free-breathing, cardiac-gated SE-cDTI acquisitions in systole and in diastole using diffusion gradients with up to second- (M2), fourth- (M4), and sixth-order (M6) motion compensation. Success rate (percentage of left ventricular voxels with (1.1 < MD < 1.9)  × 10-3mm2∕s) and diffusion measures including mean diffusivity, fractional anisotropy, helix angle, and secondary eigenvector angle were quantified and compared between both cardiac phases. Results Second-order motion compensation was sufficient for systolic cDTI, whereas fourth-order motion compensation was required for successful diastolic acquisitions in all participants. Mean diffusivity values in systole were (1.53 ± 0.06), (1.60 ± 0.07),   and   (1.62 ± 0.06)  × 10-3mm2∕s for M2, M4,   and   M6 respectively, and in diastole were (2.26 ± 0.67), (1.73 ± 0.19),   and   (1.71 ± 0.12) × 10-3mm2∕s. No significant differences in cDTI-derived parameters were observed between M4 and M6 for systolic and diastolic acquisitions. Conclusion This study demonstrates the feasibility of dual-phase SE-based cDTI techniques enabled by 300 mT/m gradients. This method will allow reliable quantification of myocardial microstructural changes between systole and diastole and represents a step towards clinical translation of dynamic cDTI.

DOI: 10.1016/j.jocmr.2026.102805 · PubMed: 42805344

Cite this article

Afzali M, Teh I, Coveney S, Mueller L, Watson I, Lwin M. Dual-phase systolic and diastolic cardiac diffusion tensor imaging using higher order motion compensation spin echo sequences at 300 mT/m gradient strength. J Cardiovasc Magn Reson. 2026. doi:10.1016/j.jocmr.2026.102805

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

More in Radiology & Imaging

Insights Imaging · Nov 20, 2026 · Research · Radiology & Imaging

Integrated system predicts disease-free survival in pancreatic cancer patients better

Eur Radiol Exp · Nov 20, 2026 · Meta-analysis · Radiology & Imaging, Oncology

MRI radiomics of pancreatic tumours: a systematic review of methodological quality and imaging reproducibility

Insights Imaging · Nov 19, 2026 · Research · Radiology & Imaging

Dual-phase photon-counting detector localized parathyroid adenomas with 67.3% to 83.7% sensitivity

EJNMMI Res · Nov 19, 2026 · Research · Radiology & Imaging

Small language model pipeline structures clinical information for dopamine transporter SPECT imaging reports

Insights Imaging · Nov 18, 2026 · Research · Radiology & Imaging

MRI detected placenta accreta spectrum with 100 percent sensitivity