ResearchClinical Chemistry · Dec 30, 2025

Evaluation of Total Bilirubin in a National Quality Program

Westenberg LEH, Lantman MVS, van Berkel M, Thelen MHM, Tintu A, Vis JY, et al
Pathology & Laboratory MedicinePediatric
Read the full paper free at the publisher ↗

Abstract

Background Accurate quantification of neonatal bilirubin remains a perennial challenge. To monitor analytical performance of total bilirubin measurements across hospitals in the Netherlands, a dedicated neonatal external quality assessment (EQA) program is in place. Methods We retrospectively analyzed Dutch EQA data from 2013 until 2025. Each year, the 81 participating laboratories received 24 pooled adult blood serum samples spiked with unconjugated bilirubin (range 100-400 µmol/L or 5.85-23.40 mg/dL). Bias from the target value determined by the reference measurement procedure (RMP) and the coefficient of variation (CV) were established for both unfactored and factored measurements. Results In total, 25 983 bilirubin measurements were analyzed. Three analytical principles were used on 12 platforms from 6 manufacturers. Diazo-based methods exhibited the lowest median bias (+0.4% in 2024), alongside spectrophotometric methods (-3.8% in 2024). Vanadate-based methods showed a bias of +19.5% in 2024 that was reduced to +3.0% for factored measurements. Substantial differences in bias were observed among manufacturers and platforms, with some exhibiting concentration dependency. Comparable patterns were seen for CVs, which generally remained below 5%. Conclusion EQA results demonstrated temporal variability in bias across all analytical principles, manufacturers, and platforms used for neonatal bilirubin assays. Bias was found to vary with bilirubin concentration. Standardization of neonatal bilirubin assays is urgently needed to ensure more reliable and accurate measurements across different systems. Meanwhile, EQA programs are important in monitoring future assay adjustments to ensure appropriate and consistent management of neonatal hyperbilirubinemia.

Subject terms

Bilirubin; Humans; Infant, Newborn; Retrospective Studies; Netherlands; Quality Assurance, Health Care

DOI: 10.1093/clinchem/hvaf144 · PubMed: 41468129

Headline, design, result, key points, caveat and citation as plain text.

Cite this article

Westenberg LEH, Lantman MVS, van Berkel M, et al. Evaluation of Total Bilirubin in a National Quality Program. Clin Chem. 2025;72(1):183-191. doi:10.1093/clinchem/hvaf144

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Pathology & Laboratory Medicine studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Research · Histopathology · May 21, 2026 · Pathology & Laboratory Medicine

Borderline breast lesions and the limits of classification: insights from the UK National Breast Screening EQA scheme

Clinical trial · Clin Chem · Apr 13, 2026 · Pathology & Laboratory Medicine

Performance, Safety, and Patient Experience of an Autonomous Robotic Phlebotomy Device: A Multicenter Trial

Research · Histopathology · Jun 19, 2026 · Pathology & Laboratory Medicine, Gastroenterology & Hepatology

Diagnostic impact and reproducibility of p53 immunohistochemistry in Barrett's oesophagus: results of the Dutch Esophageal Pathology Panel (DEPP)