Clinical trialClinical Chemistry · Apr 13, 2026

Performance, Safety, and Patient Experience of an Autonomous Robotic Phlebotomy Device: A Multicenter Trial

Giesen LFP, Roest JA, Koopman FMA, Deckers MML, Chon H, Hackeng CM, et al
Pathology & Laboratory MedicineAdult
Read the full paper free at the publisher ↗

Abstract

Background Automated phlebotomy has the potential to improve patient outcomes and address phlebotomist workforce challenges. The Autonomous Blood Drawing Optimization and Performance Testing trial (ADOPT, ClinicalTrials.gov NCT05878483) assesses performance, safety, and patient experience of an autonomous robotic phlebotomy device (ARPD) that utilizes multimodal imaging. Methods Patients were recruited from outpatient phlebotomy departments in the Netherlands. In cohort 1, analytical equivalence was assessed by comparing within-subject test results from ARPD-collected and manually collected samples using Passing-Bablok regression analysis.In cohort 2, the ARPD was assessed in routine use. The primary performance endpoint was first-stick success rate; the primary safety endpoint was the rate of adverse events. Exploratory endpoints assessed pain and patient preference. Results One hundred fifty-three patients were enrolled in cohort 1. No statistically significant differences were observed in activated partial thromboplastin time, prothrombin time, lactate dehydrogenase, aspartate transaminase, or platelet count, indicating analytical equivalence.A total of 1633 patients were enrolled in cohort 2. The overall first-stick success rate, in patients where the ARPD could identify a suitable vein, was 94.5% (95% CI, 93.3-95.5) and remained high in patients with difficult venous access (92.7%), obesity (97.4%), and age ≥65 years (93.4%). Adverse events, all mild, occurred at a rate of 0.6%. Ninety percent of patients reported far less (19%), less (32%), or similar pain (39%) compared with manual phlebotomy. Eighty-two percent indicated they would strongly prefer (16%) or prefer (31%) use of the ARPD in the future or had no preference (35%). Conclusions Automated phlebotomy using the ARPD shows a favorable performance and safety profile and may improve patient experience. ClinicalTrials.gov Registration Number: NCT05878483.

Subject terms

Humans; Phlebotomy; Female; Male; Middle Aged; Aged; Autonomous Robots; Adult; Robotics; Cohort Studies; Netherlands

DOI: 10.1093/clinchem/hvag029 · PubMed: 41967464 · Registered: NCT05878483

Headline, design, result, key points, caveat and citation as plain text.

Cite this article

Giesen LFP, Roest JA, Koopman FMA, et al. Performance, Safety, and Patient Experience of an Autonomous Robotic Phlebotomy Device: A Multicenter Trial. Clin Chem. 2026;72(8):845-856. doi:10.1093/clinchem/hvag029

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Pathology & Laboratory Medicine studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Randomized trial n = 31 · Histopathology · Oct 15, 2024 · Pathology & Laboratory Medicine, Urology

LaserSAFE fluorescence microscopy showed high accuracy for detecting positive surgical margins in prostatectomy specimens

Research · Ann Thorac Surg · Jan 20, 2026 · Cardiothoracic & Vascular Surgery, Pulmonology

Atypical Histology in Pulmonary Biopsies Obtained Using Robotic-Assisted Bronchoscopy: Natural History and Clinical Practice

Research · Clin Chem · Dec 30, 2025 · Pediatric · Pathology & Laboratory Medicine

Evaluation of Total Bilirubin in a National Quality Program

Research · Histopathology · Jun 19, 2026 · Pathology & Laboratory Medicine, Gastroenterology & Hepatology

Diagnostic impact and reproducibility of p53 immunohistochemistry in Barrett's oesophagus: results of the Dutch Esophageal Pathology Panel (DEPP)